角膜参数对不同眼压测量方法中眼压测量的影响。

Impact of corneal parameters on intraocular pressure measurements in different tonometry methods.

作者信息

Zakrzewska Aleksandra, Wiącek Marta P, Machalińska Anna

机构信息

First Department of Ophthalmology, Pomeranian Medical University, Szczecin 70-111, Poland.

出版信息

Int J Ophthalmol. 2019 Dec 18;12(12):1853-1858. doi: 10.18240/ijo.2019.12.06. eCollection 2019.

Abstract

AIM

To evaluate the impact of central corneal thickness (CCT) and corneal curvature on intraocular pressure (IOP) measurements performed by three different tonometers.

METHODS

IOP in 132 healthy eyes of 66 participants was measured using three different tonometry techniques: Goldmann applanation tonometer (GAT), Pascal dynamic contour tonometer (DCT), and ICare rebound tonometer (RT). CCT and corneal curvature were assessed.

RESULTS

In healthy eyes, DCT presents significantly higher values of IOP than GAT (17.34±3.69 and 15.27±4.06 mm Hg, <0.0001). RT measurements are significantly lower than GAT (13.56±4.33 mm Hg, <0.0001). Compared with GAT, DCT presented on average 2.51 mm Hg higher values in eyes with CCT<600 µm and 0.99 mm Hg higher results in eyes with CCT≥600 µm. The RT results were lower on average by 1.61 and 1.95 mm Hg than those obtained by GAT, respectively. Positive correlations between CCT in eyes with CCT<600 µm were detected for all IOP measurement techniques, whereas a similar relationship was not observed in eyes with thicker corneas. A correlation between IOP values and keratometry in the group with CCT<600 µm was not detected with any of the tonometry methods. In thicker corneas, a positive correlation was found for GAT and mean keratometry values (=0.369, =0.005).

CONCLUSION

The same method should always be chosen for routine IOP control, and measurements obtained by different methods cannot be compared. All analysed tonometry methods are dependent on CCT; thus, CCT should be taken into consideration for both diagnostics and monitoring.

摘要

目的

评估中央角膜厚度(CCT)和角膜曲率对三种不同眼压计测量眼压(IOP)的影响。

方法

使用三种不同的眼压测量技术对66名参与者的132只健康眼睛进行眼压测量:Goldmann压平眼压计(GAT)、Pascal动态轮廓眼压计(DCT)和ICare回弹眼压计(RT)。评估CCT和角膜曲率。

结果

在健康眼睛中,DCT测量的眼压值显著高于GAT(分别为17.34±3.69和15.27±4.06 mmHg,<0.0001)。RT测量值显著低于GAT(13.56±4.33 mmHg,<0.0001)。与GAT相比,在CCT<600 µm的眼睛中,DCT测量值平均高2.51 mmHg,在CCT≥600 µm的眼睛中高0.99 mmHg。RT测量结果分别比GAT平均低1.61和1.95 mmHg。对于所有眼压测量技术,在CCT<600 µm的眼睛中检测到CCT与眼压之间存在正相关,而在角膜较厚的眼睛中未观察到类似关系。使用任何一种眼压测量方法均未在CCT<600 µm的组中检测到眼压值与角膜曲率之间的相关性。在较厚的角膜中,发现GAT与平均角膜曲率值之间存在正相关(=0.369,=0.005)。

结论

常规眼压控制应始终选择相同的方法,不同方法获得的测量结果无法进行比较。所有分析的眼压测量方法均依赖于CCT;因此,在诊断和监测中均应考虑CCT。

相似文献

引用本文的文献

本文引用的文献

7
Comparison of current tonometry techniques in measurement of intraocular pressure.当前眼压测量技术在眼压测量中的比较。
J Curr Ophthalmol. 2016 Nov 28;29(2):92-97. doi: 10.1016/j.joco.2016.08.010. eCollection 2017 Jun.
8
Goldmann tonometer error correcting prism: clinical evaluation.戈德曼眼压计误差校正棱镜:临床评估
Clin Ophthalmol. 2017 May 3;11:835-840. doi: 10.2147/OPTH.S135272. eCollection 2017.

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验