角膜参数对不同眼压测量方法中眼压测量的影响。

Impact of corneal parameters on intraocular pressure measurements in different tonometry methods.

作者信息

Zakrzewska Aleksandra, Wiącek Marta P, Machalińska Anna

机构信息

First Department of Ophthalmology, Pomeranian Medical University, Szczecin 70-111, Poland.

出版信息

Int J Ophthalmol. 2019 Dec 18;12(12):1853-1858. doi: 10.18240/ijo.2019.12.06. eCollection 2019.

Abstract

AIM

To evaluate the impact of central corneal thickness (CCT) and corneal curvature on intraocular pressure (IOP) measurements performed by three different tonometers.

METHODS

IOP in 132 healthy eyes of 66 participants was measured using three different tonometry techniques: Goldmann applanation tonometer (GAT), Pascal dynamic contour tonometer (DCT), and ICare rebound tonometer (RT). CCT and corneal curvature were assessed.

RESULTS

In healthy eyes, DCT presents significantly higher values of IOP than GAT (17.34±3.69 and 15.27±4.06 mm Hg, <0.0001). RT measurements are significantly lower than GAT (13.56±4.33 mm Hg, <0.0001). Compared with GAT, DCT presented on average 2.51 mm Hg higher values in eyes with CCT<600 µm and 0.99 mm Hg higher results in eyes with CCT≥600 µm. The RT results were lower on average by 1.61 and 1.95 mm Hg than those obtained by GAT, respectively. Positive correlations between CCT in eyes with CCT<600 µm were detected for all IOP measurement techniques, whereas a similar relationship was not observed in eyes with thicker corneas. A correlation between IOP values and keratometry in the group with CCT<600 µm was not detected with any of the tonometry methods. In thicker corneas, a positive correlation was found for GAT and mean keratometry values (=0.369, =0.005).

CONCLUSION

The same method should always be chosen for routine IOP control, and measurements obtained by different methods cannot be compared. All analysed tonometry methods are dependent on CCT; thus, CCT should be taken into consideration for both diagnostics and monitoring.

摘要

目的

评估中央角膜厚度(CCT)和角膜曲率对三种不同眼压计测量眼压(IOP)的影响。

方法

使用三种不同的眼压测量技术对66名参与者的132只健康眼睛进行眼压测量:Goldmann压平眼压计(GAT)、Pascal动态轮廓眼压计(DCT)和ICare回弹眼压计(RT)。评估CCT和角膜曲率。

结果

在健康眼睛中,DCT测量的眼压值显著高于GAT(分别为17.34±3.69和15.27±4.06 mmHg,<0.0001)。RT测量值显著低于GAT(13.56±4.33 mmHg,<0.0001)。与GAT相比,在CCT<600 µm的眼睛中,DCT测量值平均高2.51 mmHg,在CCT≥600 µm的眼睛中高0.99 mmHg。RT测量结果分别比GAT平均低1.61和1.95 mmHg。对于所有眼压测量技术,在CCT<600 µm的眼睛中检测到CCT与眼压之间存在正相关,而在角膜较厚的眼睛中未观察到类似关系。使用任何一种眼压测量方法均未在CCT<600 µm的组中检测到眼压值与角膜曲率之间的相关性。在较厚的角膜中,发现GAT与平均角膜曲率值之间存在正相关(=0.369,=0.005)。

结论

常规眼压控制应始终选择相同的方法,不同方法获得的测量结果无法进行比较。所有分析的眼压测量方法均依赖于CCT;因此,在诊断和监测中均应考虑CCT。

相似文献

1
Impact of corneal parameters on intraocular pressure measurements in different tonometry methods.
Int J Ophthalmol. 2019 Dec 18;12(12):1853-1858. doi: 10.18240/ijo.2019.12.06. eCollection 2019.
3
Comparison of Goldmann applanation tonometry, rebound tonometry and dynamic contour tonometry in normal and glaucomatous eyes.
Int J Ophthalmol. 2015 Apr 18;8(2):299-304. doi: 10.3980/j.issn.2222-3959.2015.02.15. eCollection 2015.
4
Evaluation of Goldmann applanation tonometry, rebound tonometry and dynamic contour tonometry in keratoconus.
J Optom. 2017 Apr-Jun;10(2):117-122. doi: 10.1016/j.optom.2016.04.005. Epub 2016 Jul 9.
8
Comparison of dynamic contour tonometry with goldmann applanation tonometry.
Invest Ophthalmol Vis Sci. 2004 Sep;45(9):3118-21. doi: 10.1167/iovs.04-0018.
9
Tonometry after Intrastromal Corneal Ring Segments for Keratoconus.
Optom Vis Sci. 2017 Oct;94(10):986-992. doi: 10.1097/OPX.0000000000001120.

引用本文的文献

4
Distribution and associated factors of intraocular pressure in the older population: Tehran Geriatric Eye Study.
Int J Ophthalmol. 2023 Mar 18;16(3):418-426. doi: 10.18240/ijo.2023.03.13. eCollection 2023.
5
Comparison of intraocular pressure readings with Perkins, Tonopen, iCare 200, and iCare Home to manometry in cadaveric eyes.
Int J Ophthalmol. 2022 Dec 18;15(12):2022-2027. doi: 10.18240/ijo.2022.12.19. eCollection 2022.
6
Influence of Corneal Visualization Scheimpflug Technology Tonometry on Intraocular Pressure.
Ophthalmol Sci. 2021 Jan 13;1(1):100003. doi: 10.1016/j.xops.2021.100003. eCollection 2021 Mar.
7
Intraocular Pressure and Its Relation to Ocular Geometry: Results From the Gutenberg Health Study.
Invest Ophthalmol Vis Sci. 2022 Jan 3;63(1):40. doi: 10.1167/iovs.63.1.40.
8
Different Effects of Aging on Intraocular Pressures Measured by Three Different Tonometers.
J Clin Med. 2021 Sep 17;10(18):4202. doi: 10.3390/jcm10184202.

本文引用的文献

1
Pediatric intraocular pressure measurements: Tonometers, central corneal thickness, and anesthesia.
Surv Ophthalmol. 2019 Nov-Dec;64(6):810-825. doi: 10.1016/j.survophthal.2019.05.003. Epub 2019 May 24.
2
Comparison of rebound tonometry and non-contact airpuff tonometry to Goldmann applanation tonometry.
Ther Adv Ophthalmol. 2019 Mar 14;11:2515841419835731. doi: 10.1177/2515841419835731. eCollection 2019 Jan-Dec.
4
Agreement and repeatability of central corneal thickness measurements by four different optical devices and an ultrasound pachymeter.
Int Ophthalmol. 2019 Jul;39(7):1589-1598. doi: 10.1007/s10792-018-0983-2. Epub 2018 Jul 9.
5
Dependence of dynamic contour and Goldmann applanation tonometries on peripheral corneal thickness.
Int J Ophthalmol. 2017 Oct 18;10(10):1521-1527. doi: 10.18240/ijo.2017.10.07. eCollection 2017.
7
Comparison of current tonometry techniques in measurement of intraocular pressure.
J Curr Ophthalmol. 2016 Nov 28;29(2):92-97. doi: 10.1016/j.joco.2016.08.010. eCollection 2017 Jun.
8
Goldmann tonometer error correcting prism: clinical evaluation.
Clin Ophthalmol. 2017 May 3;11:835-840. doi: 10.2147/OPTH.S135272. eCollection 2017.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验