Loma Linda University, Loma Linda, CA 92354, USA.
Med Phys. 2012 May;39(5):2438-46. doi: 10.1118/1.3700173.
The authors present a calibration method for a prototype proton computed tomography (pCT) scanner. The accuracy of these measurements depends upon careful calibration of the energy detector used to measure the residual energy of the protons that passed through the object.
A prototype pCT scanner with a cesium iodide (CsI(Tl)) crystal calorimeter was calibrated by measuring the calorimeter response for protons of 200 and 100 MeV initial energies undergoing degradation in polystyrene plates of known thickness and relative stopping power (RSP) with respect to water. Calibration curves for the two proton energies were obtained by fitting a second-degree polynomial to the water-equivalent path length versus calorimeter response data. Using the 100 MeV calibration curve, the RSP values for a variety of tissue-equivalent materials were measured and compared to values obtained from a standard depth-dose range shift measurement using a water-tank. A cylindrical water phantom was scanned with 200 MeV protons and its RSP distribution was reconstructed using the 200 MeV calibration.
It is shown that this calibration method produces measured RSP values of various tissue-equivalent materials that agree to within 0.5% of values obtained using an established water-tank method. The mean RSP value of the water phantom reconstruction was found to be 0.995 ± 0.006.
The method presented provides a simple and reliable procedure for calibration of a pCT scanner.
作者提出了一种原型质子计算机断层扫描(pCT)扫描仪的校准方法。这些测量的准确性取决于对用于测量穿过物体的质子的剩余能量的能量探测器的仔细校准。
使用碘化铯(CsI(Tl))晶体量热计的原型 pCT 扫描仪通过测量在具有已知厚度和相对于水的相对阻止本领(RSP)的聚苯乙烯板中经历降解的 200 MeV 和 100 MeV 初始能量质子的量热计响应来进行校准。通过将水等效路径长度与量热计响应数据拟合二次多项式,获得了两种质子能量的校准曲线。使用 100 MeV 的校准曲线,测量了各种组织等效材料的 RSP 值,并将其与使用水箱进行的标准深度剂量范围移位测量获得的值进行了比较。使用 200 MeV 质子对圆柱形水模体进行扫描,并使用 200 MeV 校准重建其 RSP 分布。
结果表明,该校准方法产生的各种组织等效材料的测量 RSP 值与使用成熟的水箱方法获得的值相差在 0.5%以内。发现水模体重建的平均 RSP 值为 0.995 ± 0.006。
提出的方法为 pCT 扫描仪的校准提供了一种简单可靠的程序。