Altan Cigdem, Kaya Vedat, Basarir Berna, Celik Ugur, Azman Engin, Akar Serpil, Demirok Ahmet, Yilmaz Omer Faruk
Beyoglu Eye Training and Research Hospital, Istanbul - Turkey.
Eur J Ophthalmol. 2012 Nov-Dec;22(6):904-10. doi: 10.5301/ejo.5000150. Epub 2012 Apr 23.
The pupil diameter under low ambient illumination is diagnostically valuable for refractive surgery. The aim of study was to compare the NeurOptics® Pupillometer, Sirius®, and Ocular Wavefront Analyser® in determining scotopic pupil diameter.
A total of 96 eyes of 48 subjects were included. The scotopic pupil size was measured with 3 instruments and the measurements repeated following instillations of 1% cyclopentolate. Agreement between the instruments was assessed.
The mean measurement obtained by Sirius was significantly larger than Ocular Wavefront Analyser and NeurOptics. The Ocular Wavefront Analyser measured significantly smaller than the others. The mean cycloplegic pupillary diameters (7.73±0.70 mm with NeurOptics, 7.42±0.45 mm with Ocular Wavefront Analyser, and 8.06±0.76 mm with Sirius) were significantly different obtained by 3 instruments (p<0.001, for each; one-way analysis of variance and paired t tests).
The differences between measured pupil diameters with or without cycloplegia obtained by the NeurOptics, Sirius, and Ocular Wavefront Analyser were significant and have unacceptable levels of disagreement. These results may not be related to illumination and accommodation only, but also to measurement algorithms and technical differences of the devices.
低环境光照下的瞳孔直径对屈光手术具有诊断价值。本研究旨在比较NeurOptics瞳孔计、Sirius和眼波前分析仪在测定暗视瞳孔直径方面的表现。
共纳入48名受试者的96只眼。使用3种仪器测量暗视瞳孔大小,并在滴入1%环喷托酯后重复测量。评估仪器之间的一致性。
Sirius测得的平均数值显著大于眼波前分析仪和NeurOptics。眼波前分析仪测得的数值显著小于其他两者。3种仪器测得的平均睫状肌麻痹瞳孔直径(NeurOptics为7.73±0.70mm,眼波前分析仪为7.42±0.45mm,Sirius为8.06±0.76mm)有显著差异(每种仪器p<0.001;单因素方差分析和配对t检验)。
NeurOptics、Sirius和眼波前分析仪测得的散瞳和未散瞳瞳孔直径之间存在显著差异,且不一致程度不可接受。这些结果可能不仅与光照和调节有关,还与测量算法和设备的技术差异有关。