IEEE Trans Med Imaging. 2012 Aug;31(8):1584-92. doi: 10.1109/TMI.2012.2197218. Epub 2012 May 2.
Microwave breast imaging (using electromagnetic waves of frequencies around 1 GHz) has mostly remained at the research level for the past decade, gaining little clinical acceptance. The major hurdles limiting patient use are both at the hardware level (challenges in collecting accurate and noncorrupted data) and software level (often plagued by unrealistic reconstruction times in the tens of hours). In this paper we report improvements that address both issues. First, the hardware is able to measure signals down to levels compatible with sub-centimeter image resolution while keeping an exam time under 2 min. Second, the software overcomes the enormous time burden and produces similarly accurate images in less than 20 min. The combination of the new hardware and software allows us to produce and report here the first clinical 3-D microwave tomographic images of the breast. Two clinical examples are selected out of 400+ exams conducted at the Dartmouth Hitchcock Medical Center (Lebanon, NH). The first example demonstrates the potential usefulness of our system for breast cancer screening while the second example focuses on therapy monitoring.
微波乳房成像(使用频率在 1GHz 左右的电磁波)在过去十年中主要停留在研究阶段,临床接受程度有限。限制患者使用的主要障碍既有硬件层面的(在收集准确和未损坏的数据方面存在挑战),也有软件层面的(通常由于重建时间长达数十小时而受到困扰)。在本文中,我们报告了针对这两个问题的改进。首先,硬件能够测量到与亚厘米图像分辨率兼容的信号水平,同时保持检查时间在 2 分钟以内。其次,软件克服了巨大的时间负担,在不到 20 分钟的时间内生成同样准确的图像。新硬件和软件的结合使我们能够生成并报告这里的第一个临床三维微波层析乳房图像。从在达特茅斯-希区柯克医疗中心(新罕布什尔州黎巴嫩)进行的 400 多次检查中选择了两个临床示例。第一个示例展示了我们的系统在乳腺癌筛查方面的潜在用途,第二个示例则侧重于治疗监测。