高敏肌钙蛋白 I 检测方法的特征及其在心血管健康人群中的应用。

Characterisation of a highly sensitive troponin I assay and its application to a cardio-healthy population.

机构信息

ACT Pathology, The Canberra Hospital, Canberra, ACT, Australia.

出版信息

Clin Chem Lab Med. 2012 Feb 3;50(5):871-8. doi: 10.1515/cclm-2011-0540.

Abstract

BACKGROUND

Abbott Diagnostics have developed a new highly sensitive troponin I (hs-TnI) assay. We have assessed its analytical characteristics and applied the assay to a population of apparently cardio-healthy persons.

METHODS

We assessed imprecision, bias compared to the previous generation assay, matrix effects, and interferences and applied the assay to an apparently healthy population, deriving the 99th percentile limit of the distribution of values in reference populations for men and women separately.

RESULTS

The dynamic range of the assay was ranged from 0.5-50,000 ng/L (pg/mL). The 10% CV was at a concentration of 3.9 ng/L, and the 20% CV was at a concentration of 1.8 ng/L. The new and current version of the TnI assay were highly correlated [slope: 0.98 (95%CI:0.88-1.07), y-intercept:1.20 (95%CI:-2.35-4.75) r²=0.99]. The 99th percentile limit of the distribution of values in a reference population was different for males and females: for males 14.0 ng/L and for females 11.1 ng/L and at these concentrations the assay CV was 5.0%. TnI was detectable in nearly all patient samples from the healthy reference population (98.6%).

CONCLUSIONS

This new hs-TnI assay is able to measure to an order of magnitude lower than the current generation TnI assay from the same manufacturer. With TnI being detectable in nearly all apparently healthy subject samples this suggests that TnI presence does not always indicate cardiomyocyte necrosis.

摘要

背景

雅培诊断公司开发了一种新的高敏肌钙蛋白 I(hs-TnI)检测方法。我们评估了其分析特性,并将该检测方法应用于一组看似健康的人群。

方法

我们评估了不精密度、与上一代检测方法的偏倚、基质效应和干扰,并将该检测方法应用于一个看似健康的人群,分别为男性和女性参考人群的分布值确定第 99 百分位限值。

结果

该检测方法的动态范围为 0.5-50,000ng/L(pg/mL)。10%CV 的浓度为 3.9ng/L,20%CV 的浓度为 1.8ng/L。新的和当前版本的 TnI 检测方法高度相关[斜率:0.98(95%CI:0.88-1.07),y 截距:1.20(95%CI:-2.35-4.75),r²=0.99]。参考人群中值分布的第 99 百分位限值因男性和女性而有所不同:男性为 14.0ng/L,女性为 11.1ng/L,在这些浓度下,检测方法的 CV 为 5.0%。TnI 可在健康参考人群的几乎所有患者样本中检测到(98.6%)。

结论

与同一制造商的当前一代 TnI 检测方法相比,这种新的 hs-TnI 检测方法能够测量到数量级更低的水平。由于 TnI 在几乎所有看似健康的受试者样本中都可检测到,这表明 TnI 的存在并不总是表明心肌细胞坏死。