Suppr超能文献

一种不常见的导致骨水泥型单髁膝关节置换失败的原因:金属部件骨折。

An uncommon cause of cemented unicompartmental knee arthroplasty failure: fracture of metallic components.

机构信息

Ist Orthopedic Department, C.T.O. Hospital, Via Bignami 1, 20100, Milan, Italy,

出版信息

Knee Surg Sports Traumatol Arthrosc. 2013 Nov;21(11):2518-22. doi: 10.1007/s00167-012-2062-y. Epub 2012 May 26.

Abstract

PURPOSE

Despite good overall clinical results, unicompartmental knee replacements (UKR) are not without their problems and failures have been reported. The most common causes of UKR failure are component loosening, poor patient selection, poor surgical technique, polyethylene wear and progression of arthritis in other compartments. The purpose of this study is to present a series of atraumatic fractures of metallic components in a UKR treated in a single orthopaedic centre.

METHOD

Since 1999, 121 failed unicompartmental knee arthroplasties have been referred to our centre. In six of these, atraumatic breakage of a metal component in the cemented UKR was seen and included in this study. Pre-operative alignment, BMI and implant longevity were documented. The femoral implant failed in 4 patients and the tibial implant in a further 2.

RESULTS

All the femoral implant fractures occurred within 3 years of UKR surgery (mean: 22.2 months, SD: 10.6 months). Tibial implant breakage occurred at a mean of 8.5 years (SD: 2.4 months) following UKR. All patients were treated with conversion to a navigated total knee replacement. A primary total knee arthroplasty was used in all cases with one patient requiring a tibial component incorporating a wedge and stem following breakage of the original UKR tibial implant.

CONCLUSION

Fracture of the metallic components is a potential cause of failure of unicompartmental knee arthroplasty. In our experience, the incidence of this complication was 4.9 % of all UKR failures. Patients with a BMI greater than 30 and a progressive deterioration in limb alignment were at greater risk.

摘要

目的

尽管单髁膝关节置换术(UKR)总体临床效果良好,但并非没有问题,已有失败的报道。UKR 失败的最常见原因是部件松动、患者选择不当、手术技术不佳、聚乙烯磨损以及其他部位关节炎的进展。本研究旨在介绍在一家骨科中心治疗的 UKR 中出现的一系列金属部件非外伤性骨折。

方法

自 1999 年以来,有 121 例失败的单髁膝关节置换术被转诊至我们中心。在其中 6 例中,发现并包括了 UKR 中已粘结的金属部件出现非外伤性断裂。记录术前对线、BMI 和植入物的使用寿命。股骨植入物在 4 例患者中失败,胫骨植入物在另外 2 例中失败。

结果

所有股骨植入物骨折均发生在 UKR 手术后 3 年内(平均:22.2 个月,标准差:10.6 个月)。胫骨植入物断裂发生在 UKR 后平均 8.5 年(标准差:2.4 个月)。所有患者均采用导航全膝关节置换术进行治疗。所有病例均采用初次全膝关节置换术,1 例患者因 UKR 胫骨植入物断裂,需使用包含楔形和柄的胫骨组件。

结论

金属部件的骨折是单髁膝关节置换术失败的潜在原因。根据我们的经验,这种并发症的发生率占所有 UKR 失败的 4.9%。BMI 大于 30 和肢体对线逐渐恶化的患者风险更高。

文献AI研究员