Hearing threshold prediction with Auditory Steady State Responses and estimation of correction functions to compensate for differences with behavioral data, in adult subjects. Part 1: Audera and CHARTR EP devices.

机构信息

Department of Audiology and ENT, University of Ferrara, Ferrara, Italy.

出版信息

Med Sci Monit. 2012 Jul;18(7):MT47-53. doi: 10.12659/msm.883195.

Abstract

BACKGROUND

The objective of the study was the evaluation and comparison of hearing threshold values extrapolated from Auditory Steady State Responses, using 2 commercial systems and the estimation of correction factors applicable to the ASSR data.

MATERIAL/METHODS: One hundred ten subjects participated to the study. All subjects were initially examined with otoscopy, pure-tone audiometry and admittance. Data were acquired by 2 clinical systems the Audera (Viasys) and the CHARTR EP (ICS), using identical protocols. The acoustic stimuli consisted of single carrier frequencies at 1000, 2000 and 4000 Hz modulated at 40 Hz.

RESULTS

The data show that the threshold estimates from both devices differ significantly from the measured behavioral thresholds. The ICS device presented significantly larger mean-ASSR estimated hearing level values at the tested frequencies, implying an underestimation of the hearing threshold. Both sets of prediction errors overestimated hearing levels for the normal group. The prediction errors were in all cases greater for the Audera than for the ICS.

CONCLUSIONS

The errors encountered in the estimates of the 2 widely-used commercial devices suggest that the current ASSR protocols are not ready for a wide-range use and that significant developments in the area of threshold prediction/precision are necessary. If, on the other-hand, the ASSR predicted threshold is used on a purely consulting basis, as in hearing-aid fitting, then such errors might be acceptable in a clinical setting.

摘要

背景

本研究的目的是评估和比较使用两种商业系统从听觉稳态反应中推断出的听力阈值,并估计适用于 ASSR 数据的校正因子。

材料/方法:110 名受试者参与了这项研究。所有受试者均首先接受耳镜检查、纯音听力测试和导纳测试。数据由 2 个临床系统 Audera(Viasys)和 CHARTR EP(ICS)采集,使用相同的方案。声刺激由 1000Hz、2000Hz 和 4000Hz 的单个载波频率组成,调制频率为 40Hz。

结果

数据显示,两种设备的阈值估计值与测量的行为阈值有显著差异。ICS 设备在测试频率下呈现出显著较大的平均 ASSR 估计听力水平值,这意味着听力阈值被低估。两组预测误差均高估了正常组的听力水平。在所有情况下,Audera 的预测误差都大于 ICS。

结论

两种广泛使用的商业设备的估计中遇到的误差表明,当前的 ASSR 方案还没有准备好广泛使用,需要在阈值预测/精度方面进行重大发展。另一方面,如果 ASSR 预测的阈值仅用于咨询目的,如助听器适配,则在临床环境中,这种误差可能是可以接受的。