Wang Bo, Xiang Liangzhong, Jiang Max S, Yang Jianjun, Zhang Qizhi, Carney Paul R, Jiang Huabei
Biomed Opt Express. 2012 Jun 1;3(6):1427-32. doi: 10.1364/BOE.3.001427. Epub 2012 May 17.
A real-time three-dimensional (3D) photoacoustic imaging system was developed for epilepsy imaging in small animals. The system is based on a spherical array containing 192 transducers with a 5 MHz central frequency. The signals from the 192 transducers are amplified by 16 homemade preamplifier boards with 26 dB and multiplexed into a 64 channel data acquisition system. It can record a complete set of 3D data at a frame rate of 3.3 f/s, and the spatial resolution is about 0.2 mm. Phantom experiments were conducted to demonstrate the high imaging quality and real time imaging ability of the system. Finally, we tested the system on an acute epilepsy rat model, and the induced seizure focus was successfully detected using this system.
开发了一种用于小动物癫痫成像的实时三维(3D)光声成像系统。该系统基于一个包含192个换能器的球形阵列,中心频率为5MHz。来自192个换能器的信号由16个自制的26dB前置放大器板放大,并复用至一个64通道数据采集系统。它能够以3.3帧/秒的帧率记录完整的3D数据集,空间分辨率约为0.2mm。进行了体模实验以证明该系统的高成像质量和实时成像能力。最后,我们在急性癫痫大鼠模型上测试了该系统,并使用该系统成功检测到了诱发的癫痫病灶。