Jangda Abdul Qadir, Hussein Sherali
Radiation Oncology Department (IBN ZUHR), The Aga Khan University Hospital, Karachi, Pakistan.
J Pak Med Assoc. 2012 May;62(5):490-3.
In external beam radiation therapy (EBRT), the quality assurance (QA) of the radiation beam is crucial to the accurate delivery of the prescribed dose to the patient. One of the dosimetric parameters that require monitoring is the beam output, specified as the dose rate on the central axis under reference conditions. The aim of this project was to validate dose rate calibration of megavoltage photon beams using the International Atomic Energy Agency (IAEA)/World Health Organisation (WHO) postal audit dosimetry service. Three photon beams were audited: a 6 MV beam from the low-energy linac and 6 and 18 MV beams from a dual high-energy linac. The agreement between our stated doses and the IAEA results was within 1% for the two 6 MV beams and within 2% for the 18 MV beam. The IAEA/WHO postal audit dosimetry service provides an independent verification of dose rate calibration protocol by an international facility.
在体外放射治疗(EBRT)中,辐射束的质量保证(QA)对于向患者准确输送规定剂量至关重要。需要监测的剂量学参数之一是射束输出,其定义为参考条件下中心轴上的剂量率。本项目的目的是使用国际原子能机构(IAEA)/世界卫生组织(WHO)邮政审核剂量测定服务来验证兆伏级光子束的剂量率校准。审核了三束光子束:来自低能直线加速器的6 MV束以及来自双能高能直线加速器的6 MV和18 MV束。对于两束6 MV束,我们声明的剂量与IAEA结果之间的一致性在1%以内,对于18 MV束,一致性在2%以内。IAEA/WHO邮政审核剂量测定服务由一个国际机构对剂量率校准方案进行独立验证。