Suppr超能文献

50 年国际原子能机构/世界卫生组织放射治疗邮标剂量审核计划:从 13756 个结果中我们能学到什么?

50 Years of the IAEA/WHO postal dose audit programme for radiotherapy: what can we learn from 13756 results?

机构信息

Dosimetry and Medical Radiation Physics Section, International Atomic Energy Agency, Vienna, Austria.

出版信息

Acta Oncol. 2020 May;59(5):495-502. doi: 10.1080/0284186X.2020.1723162. Epub 2020 Feb 8.

Abstract

The IAEA/WHO postal dose audit programme has been operating since 1969 with the aim of improving the accuracy and consistency of dosimetry in radiotherapy in low-income and middle-income countries world-wide. This study summarises the 50 years' experience of audits and explores the quality of reference dosimetry in participating radiotherapy centres throughout the years. During the IAEA/WHO postal audits the dose determined from the mailed dosimeter is compared with that stated by the participant. Agreement to within ±5% is regarded acceptable whilst deviations outside ±5% limits trigger follow-up actions. Of particular interest in this study was the dependence of clinical dosimetry quality on factors related to the centre infrastructure and expertise in dosimetry of its staff. The IAEA/WHO dose audit programme noted great increase in the overall percentage of acceptable results from about 50% in its early years to 99% at present, although there is some variability of results amongst participating countries. Whereas results for younger radiotherapy machines show the agreement rate between the measured and the stated doses well above 90%, for those over 20 years old the rate dropped to <80%. Linac dosimetry was always better than Co dosimetry and multi-machine centres generally performed better than single machine centres equipped with cobalt alone. Second and subsequent participation in audits showed higher quality dosimetry than the first participation. The implementation of modern dosimetry protocols resulted in more accurate dosimetry than the use of the older protocols. Over the 50 years that the IAEA has accumulated dosimetry audit data, practices in radiotherapy centres have significantly improved. Higher quality dosimetry confirmed in audits is generally associated with better infrastructure and adequate dosimetry expertise of medical physicists in participating centres.

摘要

国际原子能机构/世界卫生组织邮政剂量审核计划自 1969 年以来一直在运行,目的是提高全球低收入和中等收入国家放射治疗剂量学的准确性和一致性。本研究总结了 50 年来的审核经验,并探讨了多年来参与放疗中心的参考剂量学质量。在国际原子能机构/世界卫生组织邮政审核中,从邮寄剂量计确定的剂量与参与者报告的剂量进行比较。如果偏差在±5%以内,则认为是可接受的,而超过±5%限制的偏差则会触发后续行动。在这项研究中,特别关注的是临床剂量学质量与中心基础设施以及工作人员剂量学专业知识相关因素的依赖性。国际原子能机构/世界卫生组织剂量审核计划注意到,总体上可接受结果的百分比从早年的约 50%大幅增加到目前的 99%,尽管参与国之间的结果存在一定的可变性。虽然对于较年轻的放疗机,测量剂量与报告剂量之间的符合率远高于 90%,但对于 20 年以上的机器,该比率下降到<80%。直线加速器剂量学总是优于钴剂量学,多台机器中心的表现通常优于仅配备钴的单台机器中心。第二次及以后的审核参与表明剂量学质量高于第一次参与。现代剂量学协议的实施比使用旧协议产生更准确的剂量学。在国际原子能机构积累剂量审核数据的 50 年中,放射治疗中心的实践有了显著改善。审核中确认的更高质量的剂量学通常与参与中心更好的基础设施和足够的剂量学专业知识相关。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验