Fontanarosa Davide, van der Meer Skadi, Bloemen-van Gurp Esther, Stroian Gabriela, Verhaegen Frank
Department of Radiation Oncology (MAASTRO), GROW - School for Oncology and Developmental Biology, Maastricht University Medical Center, Maastricht 6201 BN, The Netherlands.
Med Phys. 2012 Aug;39(8):5286-92. doi: 10.1118/1.4737571.
The purpose of this work is to assess the magnitude of speed of sound (SOS) aberrations in three-dimensional ultrasound (US) imaging systems in image guided radiotherapy. The discrepancy between the fixed SOS value of 1540 m∕s assumed by US systems in human soft tissues and its actual nonhomogeneous distribution in patients produces small but systematic errors of up to a few millimeters in the positions of scanned structures.
A correction, provided by a previously published density-based algorithm, was applied to a set of five prostate, five liver, and five breast cancer patients. The shifts of the centroids of target structures and the change in shape were evaluated.
After the correction the prostate cases showed shifts up to 3.6 mm toward the US probe, which may explain largely the reported positioning discrepancies in the literature on US systems versus other imaging modalities. Liver cases showed the largest changes in volume of the organ, up to almost 9%, and shifts of the centroids up to more than 6 mm either away or toward the US probe. Breast images showed systematic small shifts of the centroids toward the US probe with a maximum magnitude of 1.3 mm.
The applied correction in prostate and liver cancer patients shows positioning errors of several mm due to SOS aberration; the errors are smaller in breast cancer cases, but possibly becoming more important when breast tissue thickness increases.
本研究旨在评估图像引导放射治疗中三维超声(US)成像系统中声速(SOS)畸变的程度。超声系统在人体软组织中假定的固定声速值1540米∕秒与其在患者体内实际的非均匀分布之间的差异,会在扫描结构的位置上产生小但系统性的误差,误差可达几毫米。
采用先前发表的基于密度的算法对五名前列腺癌、五名肝癌和五名乳腺癌患者进行校正。评估目标结构质心的移位和形状变化。
校正后,前列腺病例显示向超声探头方向移位高达3.6毫米,这在很大程度上可以解释文献中报道的超声系统与其他成像模态之间的定位差异。肝脏病例显示器官体积变化最大,高达近9%,质心移位高达6毫米以上,远离或朝向超声探头。乳腺图像显示质心向超声探头方向有系统性的小移位,最大幅度为1.3毫米。
对前列腺癌和肝癌患者应用校正后发现,由于SOS畸变存在几毫米的定位误差;乳腺癌病例中的误差较小,但当乳腺组织厚度增加时,误差可能会变得更加显著。