Department of Radiation Oncology, University Hospital Heidelberg, Im Neuenheimer Feld 400, 69120 Heidelberg, Germany.
Z Med Phys. 2013 May;23(2):153-61. doi: 10.1016/j.zemedi.2012.08.001. Epub 2012 Aug 24.
During a helical tomotherapy a binary MLC is used for fluence modulation. The 64 pneumatically driven leaves of the MLC are either completely open or closed. The fast and frequent leaf movements result in a high demand of accuracy and stability of the MLC. This article is based on the analytical investigation of the accuracy and the stability of the MLC. Different patterns of MLC movements were generated to investigate the characteristics of the MLC. One of the considered aspects contains the friction between the leaves. The influence of variations of the compressed air on the MLC was also explored. The integrated MVCT detector of the tomotherapy system deposits the treatment data in a matrix. The detector is triggered with the linear accelerator, which is pulsed by 300Hz. The data matrix is available after the treatment. An IDL (Interactive Data Language) routine was programmed in order to analyse the matrix. The points of time, at which the leaves open (POT), and the period, in which the leaves stay open (LOT), were measured and compared with the desired values. That procedure has been repeated several times a week for approximately 6 months to investigate the stability of the MLC. Relative deviations of the LOT from -0.4% to -5.4% were measured. The friction between the leaves had no significant influence on the LOT. The available compressed air, that is used to move the leaves, depends on the number of moving leaves and also on the previous movements of the MLC. Variations of the compressed air resulted in deviations of the LOT from -1.8% to -3.7%. The measured POT deviates from the programmed POT up to -18.4ms±0.7ms. This maximal deviation correlates with a shift of the gantry angle of 0.52̊ which is negligible. The MLC has shown a stable behaviour over the 6 months. A separate consideration of the leaves showed no higher standard deviation of the LOT than ±0.7ms during the investigated time. The variation between the different leaves is much higher than the deviations of LOT caused by friction and changes of compressed air. The deviations of the LOT vary between -2.6ms and -11.0ms. The developed method is feasible in order to recognize a deterioration of the MLC performance.
在螺旋断层放射治疗中,使用二进制多叶准直器进行剂量调制。多叶准直器的 64 个气动驱动叶片要么完全打开,要么完全关闭。叶片的快速频繁运动对多叶准直器的精度和稳定性提出了很高的要求。本文基于对多叶准直器精度和稳定性的分析研究。生成了不同的多叶准直器运动模式来研究多叶准直器的特性。考虑的方面之一包含叶片之间的摩擦。还探讨了压缩空气变化对多叶准直器的影响。断层放射治疗系统的集成 MVCT 探测器将治疗数据沉积在矩阵中。探测器由以 300Hz 脉冲的线性加速器触发。治疗后即可获得数据矩阵。为了分析矩阵,用 IDL(交互式数据语言)编写了一个例程。测量并比较了叶片打开的时间点(POT)和叶片打开的时间段(LOT)与期望值。每周重复该过程几次,持续约 6 个月,以研究多叶准直器的稳定性。测量到 LOT 的相对偏差在-0.4%到-5.4%之间。叶片之间的摩擦对 LOT 没有显著影响。用于移动叶片的可用压缩空气取决于移动叶片的数量以及多叶准直器的先前运动。压缩空气的变化导致 LOT 的偏差在-1.8%到-3.7%之间。测量到的 POT 与编程的 POT 相差高达-18.4ms±0.7ms。这个最大偏差与 0.52̊的机架角度移位相关,这是可以忽略的。6 个月以来,多叶准直器一直表现出稳定的行为。对叶片的单独考虑表明,在所研究的时间内,LOT 的标准偏差没有高于±0.7ms。不同叶片之间的差异远高于 LOT 由摩擦和压缩空气变化引起的偏差。LOT 的偏差在-2.6ms 和-11.0ms 之间变化。所开发的方法能够识别多叶准直器性能的恶化。