Evaluation of three first-generation ion-selective electrode analyzers for lithium: systematic errors, frequency of random interferences, and recommendations based on comparison with flame atomic emission spectrometry.

作者信息

Okorodudu A O, Burnett R W, McComb R B, Bowers G N

机构信息

Department of Pathology, Hartford Hospital, CT 06115.

出版信息

Clin Chem. 1990 Jan;36(1):104-10.

PMID:2297898
Abstract

Ion-selective electrode analyzers for measuring lithium (Li/ISE) in serum became available in early 1987. We compared results for patients' samples from three of them vs results from flame atomic emission spectrometry (FAES). Within-run and day-to-day imprecision ranged from 0.01 to 0.03 mmol/L and 0.01 to 0.04 mmol/L, respectively. Comparing Li/ISE results (y) with the FAES results (x) gave the following equations: y = 1.063x - 0.035 for AMDEV's Lytening 2, y = 1.020x + 0.038 for NOVA's Model 11, and y = 1.030x - 0.027 for AVL's Model 985. Unexplained positive errors greater than 0.2 mmol/L were observed for two of the 90 patients' samples, but only a few additional excessively high values were seen in 3000 patients' samples run subsequently (Lytening 2). Causes of error in clinical Li/ISE measurements are still unclear; simply characterizing them as "matrix effects" does not correct the underlying analytical problem. An increase in pH from loss of CO2 gave low results on two of the three Li/ISE analyzers but did not change FAES results. Trimethylammonium bicarbonate used in a reconstitution solution caused extremely high Li/ISE results but did not change FAES results. Performance specifications to help reduce and correct these errors are recommended.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验