[Chemical sensors in medical use: ion-selective electrodes, limitations and potentials].

作者信息

Spichiger U E, Rumpf G, Haase E, Simon W

机构信息

Laboratorium für Organische Chemie, ETH Zürich.

出版信息

Schweiz Med Wochenschr. 1991 Dec 14;121(50):1875-9.

PMID:1759147
Abstract

The electrolyte concentrations of potassium and sodium, for example, are routinely analyzed in diluted serum and plasma. In contrast, electrolyte activities are quantified by ISEs directly in undiluted whole blood and plasma and promise to be highly informative. Nevertheless, some drawbacks, such as the interaction between the biological matrix and the ISE surface, limit the accuracy of measurements. A symmetrical measuring arrangement allows one to localize and quantify the known sources of errors. These are minimized by adequate techniques. The membrane composition for sodium selective electrodes is optimized in particular. The interpretation of ISE values as compared to FAES (flame atomic emission spectroscopy) is closely related to the water concentration of the samples. Comparisons between ISE and FAES results obviously allow one to evaluate the water concentration of samples, as is shown by a linear regression plot.