Ratra Vineet, Ratra Dhanashree, Gupta Muneeswar, Vaitheeswaran K
Medical Research Foundation, Sankara Nethralaya, Chennai, India.
Oman J Ophthalmol. 2012 May;5(2):97-102. doi: 10.4103/0974-620X.99372.
To determine the correlation between fundus perimetry with Micro Perimeter 1 (MP1) and conventional automated static threshold perimetry using the Humphrey Field Analyzer (HFA) in healthy individuals and in subjects with glaucoma.
In this study, we enrolled 45 eyes with glaucoma and 21 eyes of age-matched, healthy individuals. All subjects underwent complete ophthalmic examination. Differential light sensitivity was measured at 21 corresponding points in a rectangular test grid in both MP1 and HFA. Similar examination settings were used with Goldmann III stimulus, stimulus presentation time of 200 ms, and white background illumination (1.27 cd/m(2)). Statistical analysis was done with the SPSS 14 using linear regression and independent t-test.
The mean light thresholds of 21 matching points in control group with MP1 and HFA were 14.97 ± 2.64 dB and 30.90 ± 2.08 dB, respectively. In subjects with glaucoma, the mean values were MP1: 11.73 ± 4.36 dB and HFA: 27.96 ± 5.41 dB. Mean difference of light thresholds among the two instruments was 15.86 ± 3.25 dB in normal subjects (P < 0.001) and 16.22 ± 2.77 dB in glaucoma subjects (P < 0.001). Pearson correlation analysis of the HFA and MP1 results for each test point location in both cases and control subjects showed significant positive correlation (controls, r = 0.439, P = 0.047; glaucoma subjects, r = 0.812, P < 0.001). There was no difference between nasal and temporal points but a slight vertical asymmetry was observed with MP1.
There are significant and reproducible differences in the differential light threshold in MP1 and HFA in both normal and glaucoma subjects. We found a correction factor of 17.271 for comparison of MP1 with HFA. MP1 appeared to be more sensitive in predicting loss in glaucoma.
确定使用微视野计1(MP1)进行的眼底视野检查与使用汉弗莱视野分析仪(HFA)进行的传统自动静态阈值视野检查在健康个体和青光眼患者中的相关性。
在本研究中,我们纳入了45只青光眼患眼和21只年龄匹配的健康个体的眼睛。所有受试者均接受了全面的眼科检查。在MP1和HFA的矩形测试网格中的21个对应点测量了差分光敏感度。使用戈德曼III刺激、200毫秒的刺激呈现时间和白色背景照明(1.27坎德拉/平方米)进行了类似的检查设置。使用SPSS 14进行线性回归和独立t检验进行统计分析。
对照组中MP1和HFA的21个匹配点的平均光阈值分别为14.97±2.64分贝和30.90±2.08分贝。在青光眼患者中,MP1的平均值为11.73±4.36分贝,HFA的平均值为27.96±5.41分贝。两种仪器之间光阈值的平均差异在正常受试者中为15.86±3.25分贝(P<0.001),在青光眼患者中为16.22±2.77分贝(P<0.001)。对病例组和对照组中每个测试点位置上HFA和MPl结果进行的Pearson相关分析显示出显著的正相关(对照组,r=0.439,P=0.047;青光眼患者,r=0.812,P<0.001)。鼻侧和颞侧点之间没有差异,但使用MP1时观察到轻微的垂直不对称。
在正常和青光眼受试者中,MP1和HFA的差分光阈值存在显著且可重复的差异。我们发现用于比较MP1和HFA的校正因子为17.271。MP1在预测青光眼的视野缺损方面似乎更敏感。