Department of Cardiology, Erasmus MC, 3000 CA Rotterdam, the Netherlands.
Circ Arrhythm Electrophysiol. 2012 Dec;5(6):1059-63. doi: 10.1161/CIRCEP.112.975755. Epub 2012 Oct 22.
The Riata family of implantable cardioverter-defibrillator (ICD) leads is prone to a specific insulation abrasion characterized by externalization of conductor cables. The objective of this study was to determine the prevalence of externalized conductors and electrical abnormalities in Riata ICD leads by fluoroscopic screening and standard ICD interrogation.
All ICD implantation centers were contacted by the Netherlands Heart Rhythm Association Device Advisory Committee to identify all patients with an active Riata ICD lead and to perform fluoroscopic screening of the lead. In addition, the electrical integrity of the lead was assessed. As of March 1, 2012, data for 1029 active Riata leads were available; 47% of these were 8-F Riata and 53% were 7-F Riata ST. Externalized conductors were observed in 147 leads (14.3%). Proportion of externalized conductors was higher in 8-F Riata compared with 7-F Riata ST (21.4% vs 8.0%; P<0.001). Median time from implantation to detection of externalized conductors was 65.3 months. The estimated rates of externalized conductors were 6.9% and 36.6% at 5 and 8 years after implantation, respectively. Of the 147 leads with externalized conductors, 10.9% had abnormal electrical parameters vs 3.5% in nonexternalized leads (P<0.001).
The prevalence of externalized conductors in Riata leads is significantly high (14.3%) using fluoroscopic screening. The majority of externalized conductors are not detectable with standard ICD interrogation. Screening with fluoroscopy is reasonable.
植入式心律转复除颤器(ICD)的 Riata 系列导联易发生特定的绝缘磨损,表现为导体电缆外露。本研究旨在通过荧光透视筛选和标准 ICD 询问来确定 Riata ICD 导联中导体外露和电异常的发生率。
荷兰心律协会设备咨询委员会联系了所有 ICD 植入中心,以确定所有有源 Riata ICD 导联患者,并对导联进行荧光透视筛选。此外,还评估了导联的电气完整性。截至 2012 年 3 月 1 日,共有 1029 条有源 Riata 导联的数据可用;其中 47%为 8-F Riata,53%为 7-F Riata ST。在 147 条导联中观察到导体外露(14.3%)。8-F Riata 的导体外露比例高于 7-F Riata ST(21.4%比 8.0%;P<0.001)。从植入到检测到导体外露的中位时间为 65.3 个月。植入后 5 年和 8 年的导体外露估计发生率分别为 6.9%和 36.6%。在 147 条有导体外露的导联中,10.9%的电参数异常,而无导体外露的导联为 3.5%(P<0.001)。
使用荧光透视筛选,Riata 导联的导体外露发生率显著较高(14.3%)。大多数导体外露无法通过标准 ICD 询问检测到。荧光透视筛选是合理的。