NPPTL/NIOSH/CDC, Technology Research Branch, Pittsburgh, USA.
Int J Sports Med. 2013 Jun;34(6):497-501. doi: 10.1055/s-0032-1327661. Epub 2012 Nov 22.
The Zephyr BioHarness was tested to determine the accuracy of heart rate (HR) and respiratory rate (RR) measurements during 2 exercise protocols in conjunction with either a laboratory metabolic cart (Vmax) or a previously validated portable metabolic system (K4b2). In one protocol, HR and RR were measured using the BioHarness and Vmax during a graded exercise up to V˙O2max (n=12). In another protocol, HR and RR were measured using the BH and K4b2 during sustained exercise (30% and 50% V˙O2max for 20 min each) in a hot environment (30 °C, 50% relative humidity) (n=6). During the graded exercise, HR but not RR, obtained from the BioHarness was higher compared to the Vmax at baseline and 30% V˙O2max (p<0.05), but showed no significant difference at other stages with high correlation coefficients for both HR (r=0.87-0.96) and RR (r=0.90-0.99 above 30% V˙O2max). During the exercise in the heat, there were no significant differences between the BioHarness and K4b2 system. Correlation coefficients between the methods were low for HR but moderately to highly correlated (0.49-0.99) for RR. In conclusion, the BioHarness is comparable to Vmax and K4b2 over a wide range of V˙O2 during graded exercise and sustained exercise in the heat.
Zephyr 生物负荷仪在与实验室代谢箱(Vmax)或先前经过验证的便携式代谢系统(K4b2)联合进行的 2 项运动方案中,对心率(HR)和呼吸率(RR)测量的准确性进行了测试。在一项方案中,在达到最大摄氧量(V˙O2max)的递增运动过程中,使用生物负荷仪和 Vmax 测量 HR 和 RR(n=12)。在另一项方案中,在热环境(30°C,50%相对湿度)中进行 30%和 50% V˙O2max 持续运动(各 20 分钟)期间,使用 BH 和 K4b2 测量 HR 和 RR(n=6)。在递增运动中,与 Vmax 相比,来自生物负荷仪的 HR 但不是 RR 在基线和 30% V˙O2max 时更高(p<0.05),但在其他阶段没有显著差异,HR(r=0.87-0.96)和 RR(r=0.90-0.99,高于 30% V˙O2max)的相关系数均较高。在热环境下的运动中,生物负荷仪和 K4b2 系统之间没有显著差异。两种方法之间的 HR 相关系数较低,但 RR 的相关系数为中度至高度相关(0.49-0.99)。总之,在递增运动和热环境下的持续运动中,生物负荷仪在广泛的 V˙O2 范围内与 Vmax 和 K4b2 相当。