使用 Zephyr BioHarness 在递增运动和热环境下持续运动期间测量心率和呼吸率的准确性。

Measurement accuracy of heart rate and respiratory rate during graded exercise and sustained exercise in the heat using the Zephyr BioHarness.

机构信息

NPPTL/NIOSH/CDC, Technology Research Branch, Pittsburgh, USA.

出版信息

Int J Sports Med. 2013 Jun;34(6):497-501. doi: 10.1055/s-0032-1327661. Epub 2012 Nov 22.

Abstract

The Zephyr BioHarness was tested to determine the accuracy of heart rate (HR) and respiratory rate (RR) measurements during 2 exercise protocols in conjunction with either a laboratory metabolic cart (Vmax) or a previously validated portable metabolic system (K4b2). In one protocol, HR and RR were measured using the BioHarness and Vmax during a graded exercise up to V˙O2max (n=12). In another protocol, HR and RR were measured using the BH and K4b2 during sustained exercise (30% and 50% V˙O2max for 20 min each) in a hot environment (30 °C, 50% relative humidity) (n=6). During the graded exercise, HR but not RR, obtained from the BioHarness was higher compared to the Vmax at baseline and 30% V˙O2max (p<0.05), but showed no significant difference at other stages with high correlation coefficients for both HR (r=0.87-0.96) and RR (r=0.90-0.99 above 30% V˙O2max). During the exercise in the heat, there were no significant differences between the BioHarness and K4b2 system. Correlation coefficients between the methods were low for HR but moderately to highly correlated (0.49-0.99) for RR. In conclusion, the BioHarness is comparable to Vmax and K4b2 over a wide range of V˙O2 during graded exercise and sustained exercise in the heat.

摘要

Zephyr 生物负荷仪在与实验室代谢箱(Vmax)或先前经过验证的便携式代谢系统(K4b2)联合进行的 2 项运动方案中,对心率(HR)和呼吸率(RR)测量的准确性进行了测试。在一项方案中,在达到最大摄氧量(V˙O2max)的递增运动过程中,使用生物负荷仪和 Vmax 测量 HR 和 RR(n=12)。在另一项方案中,在热环境(30°C,50%相对湿度)中进行 30%和 50% V˙O2max 持续运动(各 20 分钟)期间,使用 BH 和 K4b2 测量 HR 和 RR(n=6)。在递增运动中,与 Vmax 相比,来自生物负荷仪的 HR 但不是 RR 在基线和 30% V˙O2max 时更高(p<0.05),但在其他阶段没有显著差异,HR(r=0.87-0.96)和 RR(r=0.90-0.99,高于 30% V˙O2max)的相关系数均较高。在热环境下的运动中,生物负荷仪和 K4b2 系统之间没有显著差异。两种方法之间的 HR 相关系数较低,但 RR 的相关系数为中度至高度相关(0.49-0.99)。总之,在递增运动和热环境下的持续运动中,生物负荷仪在广泛的 V˙O2 范围内与 Vmax 和 K4b2 相当。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/61f0/4620538/1028bf12c31e/nihms730660f1.jpg

相似文献

引用本文的文献

本文引用的文献

1
3
Respiration rate monitoring methods: a review.呼吸率监测方法:综述。
Pediatr Pulmonol. 2011 Jun;46(6):523-9. doi: 10.1002/ppul.21416. Epub 2011 Jan 31.
10
Validation of the COSMED K4 b2 portable metabolic system.COSMED K4 b2便携式代谢系统的验证
Int J Sports Med. 2001 May;22(4):280-4. doi: 10.1055/s-2001-13816.

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验