Suppr超能文献

植入式心律转复除颤器患者中幻肢感电击的发生率及预测因素。

Incidence and predictors of phantom shocks in implantable cardioverter defibrillator recipients.

机构信息

Department of Cardiology, Thoraxcenter Twente, Medisch Spectrum Twente, Haaksbergerstraat 55, 7513 ER, Enschede, the Netherlands,

出版信息

Neth Heart J. 2013 Apr;21(4):191-5. doi: 10.1007/s12471-012-0345-z.

Abstract

BACKGROUND

Implantable cardioverter defibrillators (ICDs) are designed to deliver shocks or antitachycardia pacing (ATP) in the event of ventricular arrhythmias. During follow-up, some ICD recipients experience the sensation of ICD discharge in the absence of an actual discharge (phantom shock). The aim of this study was to evaluate the incidence and predictors of phantom shocks in ICD recipients.

METHODS

Medical records of 629 consecutive patients with ischaemic or dilated cardiomyopathy and prior ICD implantation were studied.

RESULTS

With a median follow-up of 35 months, phantom shocks were reported by 5.1 % of ICD recipients (5.7 % in the primary prevention group and 3.7 % for the secondary prevention group; p=NS). In the combined group of primary and secondary prevention, there were no significant predictors of the occurrence of phantom shocks. However, in the primary prevention group, phantom shocks were related to a history of atrial fibrillation (p=0.03) and NYHA class <III (p=0.05). In the secondary prevention group, there were no significant predictors for phantom shocks.

CONCLUSION

Phantom shocks occur in approximately 5 % of all ICD recipients. In primary prevention patients, a relation with a history of atrial fibrillation and NYHA class <III were significant predictors for the occurrence of phantom shocks. In the secondary prevention patients, no significant predictors were found.

摘要

背景

植入式心脏复律除颤器 (ICD) 旨在针对室性心律失常时进行电击或抗心动过速起搏 (ATP)。在随访期间,一些 ICD 受者在没有实际放电的情况下会感觉到 ICD 放电(幻影电击)。本研究旨在评估 ICD 受者发生幻影电击的发生率和预测因素。

方法

研究了 629 例连续缺血性或扩张型心肌病和先前 ICD 植入的患者的病历。

结果

中位随访 35 个月,5.1%的 ICD 受者报告出现幻影电击(一级预防组为 5.7%,二级预防组为 3.7%;p=NS)。在一级和二级预防的联合组中,没有发现幻影电击发生的显著预测因素。然而,在一级预防组中,幻影电击与房颤史(p=0.03)和 NYHA 分级<III(p=0.05)有关。在二级预防组中,没有发现幻影电击的显著预测因素。

结论

大约 5%的所有 ICD 受者会出现幻影电击。在一级预防患者中,房颤史和 NYHA 分级<III 与幻影电击的发生有关。在二级预防患者中,未发现显著的预测因素。

相似文献

1
Incidence and predictors of phantom shocks in implantable cardioverter defibrillator recipients.
Neth Heart J. 2013 Apr;21(4):191-5. doi: 10.1007/s12471-012-0345-z.
2
Phantom shocks in implantable cardioverter-defibrillator recipients: impact of education level, anxiety, and depression.
Herzschrittmacherther Elektrophysiol. 2019 Sep;30(3):306-312. doi: 10.1007/s00399-019-00645-y. Epub 2019 Aug 22.
3
Psychological indices as predictors for phantom shocks in implantable cardioverter defibrillator recipients.
Pacing Clin Electrophysiol. 2014 Jun;37(6):768-73. doi: 10.1111/pace.12343. Epub 2014 Jan 23.
4
Prognostic role of left atrial enlargement in patients with implantable cardioverter defibrillators for primary prevention.
Acta Cardiol. 2022 Feb;77(1):45-50. doi: 10.1080/00015385.2020.1856491. Epub 2020 Dec 14.
9
Outcomes of defibrillator therapy in catecholaminergic polymorphic ventricular tachycardia.
Heart Rhythm. 2014 Jan;11(1):58-66. doi: 10.1016/j.hrthm.2013.10.027. Epub 2013 Oct 11.

引用本文的文献

1
2
Understanding Phantom Shocks in Implantable Cardioverter-Defibrillator Recipients.
Cureus. 2024 Jan 29;16(1):e53161. doi: 10.7759/cureus.53161. eCollection 2024 Jan.
4
Novel ventricular tachyarrhythmia detection enhancement detects undertreated life-threatening arrhythmias.
Heart Rhythm O2. 2021 Nov 18;3(1):70-78. doi: 10.1016/j.hroo.2021.11.009. eCollection 2022 Feb.
5
Phantom Shocks Associated With a Wearable Cardioverter Defibrillator.
J Med Cases. 2021 Feb;12(2):49-53. doi: 10.14740/jmc3606. Epub 2020 Dec 30.
6
One size fits all, or do we have to rethink optimal programming in implantable cardioverter-defibrillators implanted for secondary prevention?
Heart Rhythm O2. 2020 May 16;1(2):83-84. doi: 10.1016/j.hroo.2020.05.001. eCollection 2020 Jun.
7
Phantom shocks in implantable cardioverter-defibrillator recipients: impact of education level, anxiety, and depression.
Herzschrittmacherther Elektrophysiol. 2019 Sep;30(3):306-312. doi: 10.1007/s00399-019-00645-y. Epub 2019 Aug 22.

本文引用的文献

3
Improvements in 25 Years of Implantable Cardioverter Defibrillator Therapy.
Neth Heart J. 2011 Jan;19(1):24-30. doi: 10.1007/s12471-010-0047-3. Epub 2010 Dec 22.
5
Shocks, personality, and anxiety in patients with an implantable defibrillator.
Pacing Clin Electrophysiol. 2008 Jul;31(7):850-7. doi: 10.1111/j.1540-8159.2008.01099.x.
6
Psychological indices and phantom shocks in patients with ICD.
J Interv Card Electrophysiol. 2006 Apr;15(3):185-90. doi: 10.1007/s10840-006-9010-z. Epub 2006 Aug 4.
9
Amiodarone or an implantable cardioverter-defibrillator for congestive heart failure.
N Engl J Med. 2005 Jan 20;352(3):225-37. doi: 10.1056/NEJMoa043399.
10
Effectiveness of implantable defibrillators for preventing arrhythmic events and death: a meta-analysis.
J Am Coll Cardiol. 2003 May 7;41(9):1573-82. doi: 10.1016/s0735-1097(03)00253-5.