海德堡视网膜断层扫描仪 3 测量健康眼中视神经头参数的对称性:蓝山眼研究。

Symmetry of optic nerve head parameters measured by the heidelberg retina tomograph 3 in healthy eyes: the Blue Mountains Eye study.

机构信息

Centre for Vision Research, Department of Ophthalmology and Westmead Millennium Institute, University of Sydney, Australia.

出版信息

Am J Ophthalmol. 2013 Mar;155(3):518-523.e1. doi: 10.1016/j.ajo.2012.09.019. Epub 2012 Dec 4.

Abstract

PURPOSE

To assess the symmetry of optic nerve head parameters measured by the Heidelberg Retina Tomograph 3 (HRT 3) between fellow eyes in a normal elderly population.

DESIGN

Cross-sectional population-based study.

METHODS

Participants of the Blue Mountains Eye Study 10-year follow-up who did not have optic disc disease, including glaucoma, were included. Optic nerve head parameters measured by HRT 3 were compared between fellow eyes. The normal range of interocular asymmetry (larger disc minus smaller disc) was determined by the 2.5th and 97.5th percentiles.

RESULTS

A total of 1276 eligible participants had HRT scans of both eyes. HRT measurements in right eyes differed slightly in rim steepness and rim volume from those in left eyes (P < .05). The 2.5th and 97.5th percentile of interocular asymmetry limits were -0.41 and 0.45 for cup-to-disc ratio and -0.19 and 0.22 for cup-to-disc area ratio, respectively. The highest interocular correlation was found in disc area and cup area (r ranged from 0.74-0.76), whereas mean cup depth, cup volume, and retinal nerve fiber layer thickness had the poorest correlation (r ≤ 0.07). Greater optic disc area asymmetry was associated with a larger interocular difference in the other optic nerve head parameters.

CONCLUSIONS

There was minimal interocular difference and substantial interocular correlation in optic nerve head parameters measured by HRT 3. Interocular asymmetry greater than 0.2 for cup-to-disc area ratio was considered outside the normal range.

摘要

目的

评估海德堡视网膜断层扫描仪 3(HRT 3)测量的正常老年人群同眼视神经乳头参数的对称性。

设计

横断面人群基础研究。

方法

纳入蓝山眼研究 10 年随访中无视盘疾病(包括青光眼)的参与者。比较 HRT 3 测量的同眼视神经乳头参数。通过第 2.5 和 97.5 百分位数确定双眼间的正常不对称范围(较大视盘减去较小视盘)。

结果

共有 1276 名符合条件的参与者进行了双眼 HRT 扫描。右眼的 HRT 测量值在边缘陡峭度和边缘体积方面与左眼略有不同(P<0.05)。双眼间不对称限的第 2.5 和 97.5 百分位数分别为杯盘比的-0.41 和 0.45,以及杯盘面积比的-0.19 和 0.22。在盘面积和杯面积之间发现了最高的双眼间相关性(r 范围从 0.74-0.76),而平均杯深度、杯体积和视网膜神经纤维层厚度的相关性最差(r ≤ 0.07)。较大的视盘面积不对称与其他视神经乳头参数的双眼间差异较大相关。

结论

HRT 3 测量的视神经乳头参数存在最小的双眼间差异和较大的双眼间相关性。杯盘面积比的双眼间不对称大于 0.2 被认为超出了正常范围。