Clinic for Conservative Dentistry and Periodontology, School of Dental Medicine, Arnold-Heller-Str. 3, Haus 26, 24105 Kiel, Germany.
Med Oral Patol Oral Cir Bucal. 2013 Jan 1;18(1):e130-4. doi: 10.4317/medoral.18386.
The principle of the intraosseous anesthesia (IOA) relies on the perforation of the cortical plate of the bone for direct application of the local anesthetic solution into the underlying cancellous structures. During this procedure, IOA needles might accidentally come in contact with the tooth roots. The aim of the current in vitro study was to examine the consequences of this 'worst case scenario' comparing five commercially available IOA systems.
Extracted human roots were randomly perforated using five different IOA systems with a drilling time ≤5s. To simulate normal in vivo conditions, the roots were kept humid during the drilling procedure. Data was statistically evaluated using F-test (SPSS16, SPSS Inc., Chicago, USA) and the significance level was set at p ≤ 0.05.
All examined systems resulted in root perforation. Drill fractures occurred in either none 0% (Quicksleeper, Anesto, Intraflow, Stabident) or 100% (X-Tip) of the applications. Excessive heat generation, as evident by combustion odor as well as metal and tooth discoloration, appeared in 30% (Quicksleeper), 40% (Anesto), 60% (Intraflow), 90% (Stabident) and 100% (X-Tip) of all perforations.
Within the limits of in-vitro studies, the results show a potential for irreversible root damage that might be inflicted by an improper use of IOA systems.
骨内麻醉(IOA)的原理是通过穿透骨皮质板将局部麻醉溶液直接应用于下方的松质骨结构。在这个过程中,IOA 针可能会意外接触到牙根。本体外研究的目的是比较五种市售的 IOA 系统,检查这种“最坏情况”的后果。
使用五种不同的 IOA 系统对随机抽取的人牙根进行穿孔,钻孔时间≤5s。为了模拟正常的体内条件,在钻孔过程中保持根湿润。使用 F 检验(SPSS16,SPSS Inc.,芝加哥,美国)对数据进行统计评估,显著性水平设置为 p≤0.05。
所有检查的系统都导致了牙根穿孔。在 0%(Quicksleeper、Anesto、Intraflow、Stabident)或 100%(X-Tip)的应用中,出现了钻头断裂。在 30%(Quicksleeper)、40%(Anesto)、60%(Intraflow)、90%(Stabident)和 100%(X-Tip)的所有穿孔中,都出现了明显的过热现象,表现为燃烧气味以及金属和牙齿变色。
在体外研究的范围内,结果显示 IOA 系统使用不当可能导致不可逆的根损伤。