Pavlin C J, Sherar M D, Foster F S
Department of Physics, Princess Margaret Hospital, Toronto, Ontario, Canada.
Ophthalmology. 1990 Feb;97(2):244-50. doi: 10.1016/s0161-6420(90)32598-8.
The authors have developed a method of obtaining images of cross sections of the intact eye at microscopic resolution. High-frequency ultrasound transducers (100 MHz) have been developed and incorporated into imaging devices. These devices are capable of producing images to a depth of 4 mm at an axial and lateral resolution approaching 20 microns. Resolution exceeds that of current combined A- and B-scan imaging devices by a factor of approximately 10. Microscopic images of ocular structures including Schlemm's canal, cornea, iris, ciliary muscles, and retina have been produced in eye bank eyes. These studies show the feasibility of developing an apparatus to be used in the clinical setting for examining anterior structures of the eye not visible by current techniques.
作者们已经开发出一种以微观分辨率获取完整眼睛横截面图像的方法。高频超声换能器(100兆赫兹)已被开发出来并应用于成像设备中。这些设备能够在轴向和横向分辨率接近20微米的情况下生成深度达4毫米的图像。分辨率比目前的A超和B超联合成像设备高出约10倍。在眼库的眼睛中已经生成了包括施莱姆管、角膜、虹膜、睫状肌和视网膜在内的眼部结构的微观图像。这些研究表明开发一种用于临床环境中检查当前技术无法看到的眼睛前部结构的设备是可行的。