使用Lenstar和超声生物显微镜测量前房浅深度。

Measurement of shallow anterior chamber depth using Lenstar and ultrasound biomicroscopy.

作者信息

Ying Yue, Zhao Rui-Ling, Sun Ya-Nan, Sheng Qi-Lian, Kwan Ye Lin, Kong Xiang-Mei

机构信息

Eye Institute and Department of Ophthalmology, Eye & ENT Hospital, Fudan University, Shanghai 200031, China.

NHC Key Laboratory of Myopia, Fudan University, Shanghai 200031, China.

出版信息

Int J Ophthalmol. 2025 Jul 18;18(7):1276-1281. doi: 10.18240/ijo.2025.07.10. eCollection 2025.

Abstract

AIM

To study the relationship between ultrasound biomicroscopy (UBM) and Lenstar when measuring anterior chamber depth (ACD) in eyes with shallow anterior chamber, and the necessity of conducting both examinations.

METHODS

This is a retrospective observational study including 56 acute primary angle closure (APAC) eyes and 47 primary angle closure suspect (PACS) eyes with shallow ACD. ACD value measured by Lenstar and UBM were documented. The Bland-Altman plots were examined separately in all included eyes, APAC eyes and PACS eyes, for the assessment of agreement between two measurements. The agreement was compared across different population by evaluating mean difference, width of 95% limit of agreement (LoA) and the presence of proportional bias or outliers in Bland-Altman plots.

RESULTS

The average ACD in APAC eyes (1.71±0.23 mm) was significantly smaller than that in PACS eyes (1.79±0.25 mm, =0.038). Bland-Altman plots of both APAC eyes and PACS eyes showed small mean difference without the presence of proportional bias. However, compared with PACS eyes, the Bland-Altman plot of APAC eyes had wider 95% LoA and more outliers outside the 95% LoA.

CONCLUSION

Despite the small ACD values, the two methods in measuring ACD, Lenstar and UBM, demonstrate substantial consistency when measuring PACS eyes. However, for the APAC population, the agreement might be limited, and thus the measured values should be interpreted with caution.

摘要

目的

研究超声生物显微镜(UBM)与Lenstar在测量浅前房眼的前房深度(ACD)时的关系,以及同时进行这两项检查的必要性。

方法

这是一项回顾性观察研究,纳入了56只急性原发性闭角型青光眼(APAC)眼和47只原发性闭角型青光眼可疑(PACS)眼且前房浅。记录Lenstar和UBM测量的ACD值。分别在所有纳入的眼、APAC眼和PACS眼中检查Bland-Altman图,以评估两种测量方法之间的一致性。通过评估平均差异、95%一致性界限(LoA)的宽度以及Bland-Altman图中比例偏差或异常值的存在情况,比较不同人群之间的一致性。

结果

APAC眼的平均ACD(1.71±0.23mm)显著小于PACS眼(1.79±0.25mm,P=0.038)。APAC眼和PACS眼的Bland-Altman图均显示平均差异较小且无比例偏差。然而,与PACS眼相比,APAC眼的Bland-Altman图95%LoA更宽,且95%LoA外的异常值更多。

结论

尽管ACD值较小,但Lenstar和UBM这两种测量ACD值方法在测量PACS眼时显示出较高的一致性。然而,对于APAC人群,一致性可能有限,因此对测量值的解释应谨慎。

相似文献

本文引用的文献

9
Anterior chamber depth measurement using Pentacam and Biograph in children.Pentacam 和 Biograph 在儿童中前房深度测量。
Clin Exp Optom. 2022 Aug;105(6):582-586. doi: 10.1080/08164622.2021.1971043. Epub 2021 Sep 6.
10
Glaucoma.青光眼。
Med Clin North Am. 2021 May;105(3):493-510. doi: 10.1016/j.mcna.2021.01.004. Epub 2021 Apr 2.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验