Swiss Laboratory for Doping Analyses, University Center of Legal Medicine, West Switzerland, Chemin des Croisettes 22, 1066 Epalinges, Switzerland.
J Chromatogr A. 2013 May 31;1292:142-50. doi: 10.1016/j.chroma.2012.12.008. Epub 2012 Dec 19.
Clenbuterol is a β2 agonist agent with anabolic properties given by the increase in the muscular mass in parallel to the decrease of the body fat. For this reason, the use of clenbuterol is forbidden by the World Anti-Doping Agency (WADA) in the practice of sport. This compound is of particular interest for anti-doping authorities and WADA-accredited laboratories due to the recent reporting of risk of unintentional doping following the eating of meat contaminated with traces of clenbuterol in some countries. In this work, the development and the validation of an ultra-high pressure liquid chromatography coupled to electrospray ionization tandem mass spectrometry (UHPLC-ESI-MS/MS) method for the quantification of clenbuterol in human urine is described. The analyte was extracted from urine samples by liquid-liquid extraction (LLE) in basic conditions using tert butyl-methyl ether (TBME) and analyzed by UHPLC-MS/MS with a linear gradient of acetonitrile in 9min only. The simple and rapid method presented here was validated in compliance with authority guidelines and showed a limit of quantification at 5pg/mL and a linearity range from 5pg/mL to 300pg/mL. Good trueness (85.8-105%), repeatability (5.7-10.6% RSD) and intermediate precision (5.9-14.9% RSD) results were obtained. The method was then applied to real samples from eighteen volunteers collecting urines after single oral doses administration (1, 5 and 10μg) of clenbuterol-enriched yogurts.
克伦特罗是一种具有合成代谢特性的β2 激动剂,可增加肌肉质量,同时减少体脂。出于这个原因,世界反兴奋剂机构(WADA)禁止在运动实践中使用克伦特罗。由于最近一些国家报道称,食用含有痕量克伦特罗的污染肉类会导致无意兴奋剂,这种化合物引起了反兴奋剂机构和 WADA 认可实验室的特别关注。在这项工作中,描述了一种超高效液相色谱-电喷雾串联质谱(UHPLC-ESI-MS/MS)方法的开发和验证,用于定量人尿液中的克伦特罗。在碱性条件下,通过液-液萃取(LLE)用叔丁基-甲基醚(TBME)从尿液样品中提取分析物,并在 9 分钟内通过 UHPLC-MS/MS 进行线性梯度洗脱。这里提出的简单快速方法符合权威指南的要求,并显示出 5pg/mL 的定量限和 5pg/mL 至 300pg/mL 的线性范围。获得了良好的准确度(85.8-105%)、重复性(5.7-10.6%RSD)和中间精密度(5.9-14.9%RSD)结果。然后将该方法应用于十八名志愿者的真实尿液样本,这些志愿者在单次口服剂量(1、5 和 10μg)克伦特罗浓缩酸奶后采集尿液。