一种单片非球面疏水性丙烯酸酯人工晶状体设计对其中心定位和旋转的影响。

Effect of a single-piece aspheric hydrophobic acrylic intraocular lens design on centration and rotation.

机构信息

Ophthalmology Department, Ain Shams University and National Eye Hospital, Cairo, Egypt.

出版信息

J Cataract Refract Surg. 2013 Mar;39(3):408-13. doi: 10.1016/j.jcrs.2012.09.020. Epub 2013 Jan 11.

Abstract

PURPOSE

To assess the rotation and centration stability of the new design features of a 1-piece aspheric hydrophobic acrylic intraocular lens (IOL).

SETTING

National Eye Hospital, Cairo, Egypt.

DESIGN

Randomized clinical trial.

METHODS

Patients in this study had implantation of a 1-piece Tecnis 1 ZCB00 IOL or a 3-piece Sensar AR40e IOL. Decentration and rotation were recorded immediately postoperatively, and 2 to 3 days and 8 weeks postoperatively. This was done using 2 slitlamp eyepieces and referring to constant corneal landmarks (ie, the limbus and the phaco incision).

RESULTS

Thirty-two 1-piece IOLs and 30 3-piece IOLs were implanted. Although the 1-piece IOLs moved toward the corneal center more than the 3-piece IOLs between 3 days and 8 weeks postoperatively (P=.054), the difference was not statistically significant between 0 day and 3 days or during the entire follow-up (P=.546 and P=.367, respectively). There were no statistically significant differences in the horizontal or vertical components of this movement between the 2 groups (P=.883 and P=.636, respectively). The mean rotation was 2.5 degrees ± 9.1 (SD) in the 1-piece IOL group and 0 ± 6.5 degrees in the 3-piece IOL group, with no statistically significant difference between the groups (P=.521).

CONCLUSION

The new 1-piece design showed relatively good centration and stability in the horizontal, vertical, and rotational aspects.

摘要

目的

评估一款一体式非球面疏水性丙烯酸人工晶状体(IOL)的新设计特征的旋转和中心稳定性。

地点

埃及开罗国家眼科医院。

设计

随机临床试验。

方法

本研究中的患者接受了一款一体式 Tecnis 1 ZCB00 IOL 或 3 件套 Sensar AR40e IOL 的植入。术后即刻、术后 2 至 3 天和 8 周记录偏心和旋转。这是通过使用 2 个裂隙灯目镜并参考恒定的角膜标记物(即角膜缘和白内障切口)来完成的。

结果

植入了 32 个一体式 IOL 和 30 个 3 件套 IOL。尽管一体式 IOL 在术后 3 天至 8 周内比 3 件套 IOL 更向角膜中心移动(P=.054),但在 0 天至 3 天或整个随访期间,差异无统计学意义(P=.546 和 P=.367,分别)。两组之间这种运动的水平或垂直分量均无统计学差异(P=.883 和 P=.636,分别)。一体式 IOL 组的平均旋转为 2.5 度±9.1(标准差),3 件套 IOL 组为 0 度±6.5 度,两组间无统计学差异(P=.521)。

结论

新的一体式设计在水平、垂直和旋转方面表现出相对较好的定位和稳定性。