Ophthalmology Department, Ain Shams University and National Eye Hospital, Cairo, Egypt.
J Cataract Refract Surg. 2013 Mar;39(3):408-13. doi: 10.1016/j.jcrs.2012.09.020. Epub 2013 Jan 11.
To assess the rotation and centration stability of the new design features of a 1-piece aspheric hydrophobic acrylic intraocular lens (IOL).
National Eye Hospital, Cairo, Egypt.
Randomized clinical trial.
Patients in this study had implantation of a 1-piece Tecnis 1 ZCB00 IOL or a 3-piece Sensar AR40e IOL. Decentration and rotation were recorded immediately postoperatively, and 2 to 3 days and 8 weeks postoperatively. This was done using 2 slitlamp eyepieces and referring to constant corneal landmarks (ie, the limbus and the phaco incision).
Thirty-two 1-piece IOLs and 30 3-piece IOLs were implanted. Although the 1-piece IOLs moved toward the corneal center more than the 3-piece IOLs between 3 days and 8 weeks postoperatively (P=.054), the difference was not statistically significant between 0 day and 3 days or during the entire follow-up (P=.546 and P=.367, respectively). There were no statistically significant differences in the horizontal or vertical components of this movement between the 2 groups (P=.883 and P=.636, respectively). The mean rotation was 2.5 degrees ± 9.1 (SD) in the 1-piece IOL group and 0 ± 6.5 degrees in the 3-piece IOL group, with no statistically significant difference between the groups (P=.521).
The new 1-piece design showed relatively good centration and stability in the horizontal, vertical, and rotational aspects.
评估一款一体式非球面疏水性丙烯酸人工晶状体(IOL)的新设计特征的旋转和中心稳定性。
埃及开罗国家眼科医院。
随机临床试验。
本研究中的患者接受了一款一体式 Tecnis 1 ZCB00 IOL 或 3 件套 Sensar AR40e IOL 的植入。术后即刻、术后 2 至 3 天和 8 周记录偏心和旋转。这是通过使用 2 个裂隙灯目镜并参考恒定的角膜标记物(即角膜缘和白内障切口)来完成的。
植入了 32 个一体式 IOL 和 30 个 3 件套 IOL。尽管一体式 IOL 在术后 3 天至 8 周内比 3 件套 IOL 更向角膜中心移动(P=.054),但在 0 天至 3 天或整个随访期间,差异无统计学意义(P=.546 和 P=.367,分别)。两组之间这种运动的水平或垂直分量均无统计学差异(P=.883 和 P=.636,分别)。一体式 IOL 组的平均旋转为 2.5 度±9.1(标准差),3 件套 IOL 组为 0 度±6.5 度,两组间无统计学差异(P=.521)。
新的一体式设计在水平、垂直和旋转方面表现出相对较好的定位和稳定性。