Instituto Oftalmológico de Alicante, Vissum Corporation, Alicante, Spain.
Cornea. 2013 Jun;32(6):791-8. doi: 10.1097/ICO.0b013e31827bf898.
To evaluate the correlation of the magnitude of corneal toricity and power vector components of both corneal surfaces measured with a Scheimpflug photography-based system.
A total of 117 healthy normal eyes of 117 subjects selected randomly with ages ranging from 7 to 80 years were included. All eyes received an anterior segment and corneal analysis with the Sirius system (CSO) evaluating the anterior and posterior mean toricity for 3 and 7 mm (aAST and pAST). The vector components J0 and J45 as well as the overall strength blur (B) were calculated for each keratometric measurement using the procedure defined by Thibos and Horner.
The coefficient of correlation between aAST and pAST was 0.52 and 0.62 and the mean anteroposterior ratio for toricity was 0.46 ± 0.39 and 0.57 ± 0.75 for 3 and 7 mm, respectively. These ratios correlated significantly with aAST, anterior corneal J0, and manifest refraction J0 (r ≥ 0.39, P < 0.01). The coefficient of correlation was 0.69 and 0.81 between anterior and posterior J0 for 3 and 7 mm, respectively. For J45, the coefficients were 0.62 and 0.71, respectively. The linear regression analysis revealed that the pAST and power vectors could be predicted from the anterior corneal data (R2 ≥ 0.40, P < 0.01).
The toricity and astigmatic power vector components of the posterior corneal surface in the human healthy eye are related to those of the anterior and therefore can be predicted consistently from the anterior toricity and astigmatic power vectors.
评估基于 Scheimpflug 摄影系统测量的两个角膜表面的角膜旋度幅度和力矢量分量之间的相关性。
共纳入 117 名年龄在 7 至 80 岁之间的健康正常眼,均进行了 Sirius 系统(CSO)的眼前节和角膜分析,评估了 3mm 和 7mm 的前、后平均旋度(aAST 和 pAST)。使用 Thibos 和 Horner 定义的程序,为每个角膜曲率测量计算了 J0 和 J45 矢量分量以及整体强度彗差(B)。
aAST 和 pAST 之间的相关系数分别为 0.52 和 0.62,旋度的前后比均值分别为 0.46 ± 0.39 和 0.57 ± 0.75,分别用于 3mm 和 7mm。这些比率与 aAST、前角膜 J0 和显性折射 J0 显著相关(r≥0.39,P<0.01)。3mm 和 7mm 的前、后 J0 之间的相关系数分别为 0.69 和 0.81。对于 J45,系数分别为 0.62 和 0.71。线性回归分析表明,pAST 和力矢量可以从前角膜数据预测(R2≥0.40,P<0.01)。
人眼健康后角膜的旋度和散光力矢量分量与前角膜相关,因此可以从前角膜的旋度和散光力矢量一致地预测。