基于新 Scheimpflug 摄影系统的角膜地形图仪测量角膜锥状患者的同体重复性。

Intrasubject repeatability in keratoconus-eye measurements obtained with a new Scheimpflug photography-based system.

机构信息

Instituto Oftalmológico de Alicante, Vissum Corporation, Alicante, Spain.

出版信息

J Cataract Refract Surg. 2013 Feb;39(2):211-8. doi: 10.1016/j.jcrs.2012.10.033. Epub 2012 Dec 3.

Abstract

PURPOSE

To evaluate in keratoconus eyes the intrasubject repeatability of anterior and posterior corneal curvature and of other anterior segment anatomic measurements obtained with a new topography system combining Scheimpflug-photography and Placido-disk technology.

SETTING

Vissum Corporation, Alicante, Spain.

DESIGN

Evaluation of technology.

METHODS

All keratoconus eyes had a comprehensive ophthalmologic examination including analysis with the Sirius system. Three consecutive measurements were obtained to assess the intrasubject repeatability of the following parameters: anterior and posterior corneal curvature and shape factor, white-to-white (WTW) corneal diameter, central and minimum corneal thickness, and anterior chamber depth (ACD). The within-subject standard deviation (S(w)) and intraclass correlation coefficient (ICC) were calculated.

RESULTS

This study comprised 61 eyes of 61 patients ranging in age from 14 to 64 years. For anterior and posterior corneal curvatures and power vector components, the S(w) was 0.29 mm or less in all cases. The ICC was above 0.990 in all cases except the flattest curvature of the posterior corneal surface at 3.0 mm, which was 0.840 (moderate agreement), and the posterior power vector J0, which was 0.665 (poor agreement), 0.752, and 0.758 (moderate agreement) for 3.0 mm, 5.0 mm, and 7.0 mm, respectively. In shape factor measurements, the S(w) was 0.12 or less in all cases and the ICC ranged between 0.989 and 0.999. Pachymetry, ACD, and WTW had ICC values very close to 1.

CONCLUSION

The new topography system provided repeatable measurements of corneal shape and other anatomic parameters in eyes with keratoconus.

FINANCIAL DISCLOSURE

No author has a financial or proprietary interest in any material or method mentioned.

摘要

目的

评估角膜地形图系统结合 Scheimpflug 摄影和 Placido 盘技术在圆锥角膜眼中对前、后角膜曲率和其他眼前节解剖测量的同体重复性。

设置

西班牙阿利坎特的 Vissum 公司。

设计

技术评估。

方法

所有圆锥角膜眼均进行全面眼科检查,包括 Sirius 系统分析。获得 3 次连续测量,以评估以下参数的同体重复性:前、后角膜曲率和形状因子,角膜白到白(WTW)直径,中央和最小角膜厚度,以及前房深度(ACD)。计算个体内标准差(S(w))和组内相关系数(ICC)。

结果

本研究共纳入 61 例 61 只眼,年龄 14 至 64 岁。对于前、后角膜曲率和力向量分量,所有情况下 S(w)均小于等于 0.29 毫米。除最平坦的后角膜表面曲率(3.0 毫米)为 0.840(中度一致)、后力向量 J0 为 0.665(差)外,所有情况下 ICC 均大于 0.990,后力向量 J0 在 3.0 毫米、5.0 毫米和 7.0 毫米时分别为 0.752 和 0.758(中度一致)。在形状因子测量中,所有情况下 S(w)均小于等于 0.12,ICC 范围在 0.989 至 0.999 之间。角膜厚度、ACD 和 WTW 的 ICC 值非常接近 1。

结论

新型角膜地形图系统在圆锥角膜眼中提供了可重复的角膜形状和其他解剖参数测量。

财务披露

没有作者在任何提到的材料或方法上有财务或所有权利益。