Salemyr Mats, Muren Olle, Ahl Torbjörn, Bodén Henrik, Chammout Ghazi, Stark André, Sköldenberg Olof
Division of Orthopaedics, Department of Clinical Sciences at Danderyd Hospital, Karolinska Institutet, Stockholm, Sweden.
Int Orthop. 2015 Aug;39(8):1499-505. doi: 10.1007/s00264-015-2680-3. Epub 2015 Jan 29.
We hypothesized that a vitamin E-treated polyethylene articulating liner had lower femoral head penetration than a conventional liner.
Fifty-one patients with primary hip osteoarthritis were randomized to an uncemented cup with either vitamin E-treated highly cross-linked polyethylene (HXLPE) (vitamin-e group, n = 25) or a standard HXLPE liner (control group, n = 26) and followed for two years after surgery. The outcome variable was polyethylene (PE) linear head penetration, measured with radiostereometry (RSA).
The head penetration in x- (medial/lateral) and y- (vertical) axes was significantly lower for the vitamin E-treated liner up to two years with a mean difference between the groups of 0.10 mm (95 % CI 0.04-0.18) and 0.08 mm (95 % CI 0.01-0.15), respectively. The initial total penetration was lower for the vitamin-E group but did not reach statistical significance at two years with a mean difference between the groups of 0.07 mm (95 % CI -0.001 to 0.15, p = 0.09). We found no differences in clinical outcome between the groups.
In this prospective, randomized, controlled trial on a new vitamin-E diffused HXLPE liner we found, compared to a conventional HXLPE liner, lower initial head penetration and lower superior and medial wear. A longer follow-up than two years is necessary to demonstrate lower overall head penetration in vitamin-E liners compared to contemporary highly cross-linked PE.