Butlin Mark, Qasem Ahmad, Avolio Alberto P
Australian School of Advanced Medicine, Macquarie University, Sydney, NSW, Australia.
Annu Int Conf IEEE Eng Med Biol Soc. 2012;2012:2591-4. doi: 10.1109/EMBC.2012.6346494.
There is increasing interest in non-invasive estimation of central aortic waveform parameters in the clinical setting. However, controversy has arisen around radial tonometric based systems due to the requirement of a trained operator or lack of ease of use, especially in the clinical environment. A recently developed device utilizes a novel algorithm for brachial cuff based assessment of aortic pressure values and waveform (SphygmoCor XCEL, AtCor Medical). The cuff was inflated to 10 mmHg below an individual's diastolic blood pressure and the brachial volume displacement waveform recorded. The aortic waveform was derived using proprietary digital signal processing and transfer function applied to the recorded waveform. The aortic waveform was also estimated using a validated technique (radial tonometry based assessment, SphygmoCor, AtCor Medical). Measurements were taken in triplicate with each device in 30 people (17 female) aged 22 to 79 years of age. An average for each device for each individual was calculated, and the results from the two devices were compared using regression and Bland-Altman analysis. A high correlation was found between the devices for measures of aortic systolic (R(2)=0.99) and diastolic (R(2)=0.98) pressure. Augmentation index and subendocardial viability ratio both had a between device R(2) value of 0.82. The difference between devices for measured aortic systolic pressure was 0.5±1.8 mmHg, and for augmentation index, 1.8±7.0%. The brachial cuff based approach, with an individualized sub-diastolic cuff pressure, provides an operator independent method of assessing not only systolic pressure, but also aortic waveform features, comparable to existing validated tonometric-based methods.
在临床环境中,对中心主动脉波形参数的无创估计越来越受到关注。然而,基于桡动脉压计的系统引发了争议,因为需要训练有素的操作人员,或者使用起来不够便捷,尤其是在临床环境中。最近开发的一种设备利用一种新颖的算法,通过肱动脉袖带评估主动脉压力值和波形(SphygmoCor XCEL,AtCor Medical公司)。将袖带充气至低于个体舒张压10 mmHg,并记录肱动脉容积位移波形。使用专有的数字信号处理和应用于记录波形的传递函数得出主动脉波形。还使用一种经过验证的技术(基于桡动脉压计的评估,SphygmoCor,AtCor Medical公司)来估计主动脉波形。对30名年龄在22至79岁之间的人(17名女性),分别使用这两种设备进行了三次测量。计算出每个人每种设备的平均值,并使用回归分析和布兰德-奥特曼分析比较两种设备的结果。发现两种设备在主动脉收缩压(R(2)=0.99)和舒张压(R(2)=0.98)测量值之间具有高度相关性。增强指数和心内膜下存活比值在两种设备之间的R(2)值均为0.82。两种设备测量的主动脉收缩压差值为0.5±1.8 mmHg,增强指数差值为1.8±7.0%。基于肱动脉袖带的方法,采用个体化的舒张压低值袖带压力,提供了一种独立于操作人员的方法,不仅可以评估收缩压,还可以评估主动脉波形特征,与现有的经过验证的基于压计的方法相当。