Whitacre M M, Manoukian N, Mainster M A
Department of Ophthalmology, Kansas University Medical Center, Kansas City 66103.
Br J Ophthalmol. 1990 Apr;74(4):233-4. doi: 10.1136/bjo.74.4.233.
A new argon indirect ophthalmoscopic photocoagulator is presented which uses a red helium-neon laser aiming beam. An interference filter protects the operator from green or blue-green treatment beam reflections without impairing visualisation of the aiming beam or significant retinal anatomy. The protective filter is fixed in place, eliminating the weight, noise, and potential failure of mechanically switched filters. The red aiming beam has a negligible potential for producing photochemical retinal damage in the patient or operator.
介绍了一种新型氩间接检眼镜光凝器,它使用红色氦氖激光瞄准光束。干涉滤光片可保护操作人员免受绿色或蓝绿色治疗光束反射的影响,同时又不影响瞄准光束或重要视网膜结构的可视化。保护滤光片固定在位,消除了机械切换滤光片的重量、噪音和潜在故障。红色瞄准光束对患者或操作人员产生光化学性视网膜损伤的可能性可忽略不计。