Gait Posture. 2013 Jul;38(3):544-8. doi: 10.1016/j.gaitpost.2013.01.028. Epub 2013 Feb 28.
Plantar pressure measurement is common practice in many research and clinical protocols. While the accuracy of some plantar pressure measuring devices and methods for ensuring consistency in data collection on plantar pressure measuring devices have been reported, the reliability of different devices when testing the same individuals is not known. This study calculated intra-mat, intra-manufacturer, and inter-manufacturer reliability of plantar pressure parameters as well as the number of plantar pressure trials needed to reach a stable estimate of the mean for an individual. Twenty-two healthy adults completed ten walking trials across each of two Novel emed-x(®) and two Tekscan MatScan(®) plantar pressure measuring devices in a single visit. Intraclass correlation (ICC) was used to describe the agreement between values measured by different devices. All intra-platform reliability correlations were greater than 0.70. All inter-emed-x(®) reliability correlations were greater than 0.70. Inter-MatScan(®) reliability correlations were greater than 0.70 in 31 and 52 of 56 parameters when looking at a 10-trial average and a 5-trial average, respectively. Inter-manufacturer reliability including all four devices was greater than 0.70 for 52 and 56 of 56 parameters when looking at a 10-trial average and a 5-trial average, respectively. All parameters reached a value within 90% of an unbiased estimate of the mean within five trials. Overall, reliability results are encouraging for investigators and clinicians who may have plantar pressure data sets that include data collected on different devices.
足底压力测量在许多研究和临床方案中是常见的做法。虽然已经报道了一些足底压力测量设备的准确性和确保足底压力测量设备数据收集一致性的方法,但不同设备在测试同一个体时的可靠性尚不清楚。本研究计算了足底压力参数的内-设备、内-制造商和间-制造商可靠性,以及达到个体平均稳定估计所需的足底压力测试次数。22 名健康成年人在单次就诊中完成了在两个 Novel emed-x(®)和两个 Tekscan MatScan(®)足底压力测量设备上各进行 10 次步行测试。使用组内相关系数(ICC)来描述不同设备测量值之间的一致性。所有内平台可靠性相关系数均大于 0.70。所有 emed-x(®)间可靠性相关系数均大于 0.70。当分别观察 10 次测试平均值和 5 次测试平均值时,56 个参数中的 31 个和 52 个 MatScan(®)间可靠性相关系数大于 0.70。当分别观察 10 次测试平均值和 5 次测试平均值时,包括所有四个设备的制造商间可靠性在 56 个参数中的 52 个和 56 个参数中大于 0.70。所有参数在五次测试内均达到无偏估计平均值的 90%以内。总的来说,对于可能有包括不同设备收集的数据的足底压力数据集的研究人员和临床医生来说,可靠性结果令人鼓舞。