Reliability of pressure measurements: the EM ED F system.

作者信息

Hughes J, Pratt L, Linge K, Clark P, Klenerman L

机构信息

University Department of Orthopaedic and Accident Surgery, Royal Liverpool Hospital, Liverpool, UK; Section of Medical Statistics, Clinical Research Centre, Watford Road, Harrow, UK.

出版信息

Clin Biomech (Bristol). 1991 Feb;6(1):14-8. doi: 10.1016/0268-0033(91)90036-P.

Abstract

The reliability of foot pressure measurement using the EMED F system has been tested by recording 25 walks with 10 volunteer subjects at three different speeds. Coefficients of reliability calculated for 1-25 walks showed that a good level of reliability was achieved using one measurement for most force/pressure variables but when the mean result of three or more walks was used reliability was excellent. Measurements related to time were more variable than the total force, the peak pressure and the area. The total force and peak pressure were also shown to increase with increasing speed but this was not true for all sites on the foot.

摘要

通过记录10名志愿者在三种不同速度下的25次行走,对使用EMED F系统进行足部压力测量的可靠性进行了测试。对第1 - 25次行走计算出的可靠性系数表明,对于大多数力/压力变量,使用一次测量就能达到较好的可靠性水平,但当使用三次或更多次行走的平均结果时,可靠性极佳。与时间相关的测量比总力、峰值压力和面积的变化更大。总力和峰值压力也显示出随着速度增加而增加,但并非足部所有部位都是如此。