光学相干断层扫描测量泪膜新月形:与泪液分泌试验和泪膜破裂时间的相关性。

Optical coherence tomography for measuring the tear film meniscus: correlation with schirmer test and tear-film breakup time.

机构信息

Faculty of Medicine, Adana Teaching & Medical Research Center, Baskent University, Adana, Turkey.

出版信息

Curr Eye Res. 2013 Jul;38(7):736-42. doi: 10.3109/02713683.2013.774422. Epub 2013 Mar 14.

Abstract

PURPOSE

To compare tear meniscus height (TMH) and area (TMA) values obtained by optical coherence tomography (OCT) with the Schirmer test and tear-film break-up time (TBUT) values.

MATERIALS AND METHODS

In this prospective cross-sectional study, the right eyes of 300 consecutive patients were studied. All patients underwent routine ophthalmologic examination. Schirmer and TBUT measurements were taken, and tear-film OCT images were obtained. Using OCT, the TMH (i.e. the line connecting the intersection of the meniscus with the cornea and eyelid) and cross-sectional TMA were calculated. Patients were divided into Groups 1 (Schirmer values ≤5 mm) and 2 (Schirmer values >5 mm). Data were analyzed using SPSS 13.0 software. The Mann-Whitney U-test was used for the comparison of groups. Correlations between tear parameters were analyzed.

RESULTS

There were no significant differences in age, logMAR, or IOP values (p = 0.480, 0.077 and 0.146, respectively) between the two groups. Mean TBUTs were 5.1 ± 2.9 and 9.5 ± 4.2 s for Groups 1 and 2, respectively (p < 0.001). Mean TMHs were 237.9 ± 108.9 and 324.3 ± 158.9 μm for Groups 1 and 2, respectively (p < 0.001). Mean TMAs were significantly lower in Group 1 compared to Group 2 (p < 0.001; 0.027 ± 0.028 versus 0.055 ± 0.059 mm(2), respectively). Correlations between Schirmer values and TMH, and between Schirmer values and TMA, were significant (p = 0.001 and <0.001, respectively).

CONCLUSIONS

OCT values were significantly lower in patients with Schirmer values of <5 mm. Tear meniscus measurements obtained by OCT are reliable for establishing a diagnosis of dry eye.

摘要

目的

比较光学相干断层扫描(OCT)测量的泪膜半月板高度(TMH)和面积(TMA)与 Schirmer 测试和泪膜破裂时间(TBUT)值。

材料与方法

本前瞻性横断面研究纳入了 300 例连续患者的右眼。所有患者均接受了常规眼科检查。进行了 Schirmer 和 TBUT 测量,并获得了泪膜 OCT 图像。使用 OCT 计算 TMH(即连接半月板与角膜和眼睑交点的线)和横截面积 TMA。患者分为两组:1 组(Schirmer 值≤5mm)和 2 组(Schirmer 值>5mm)。使用 SPSS 13.0 软件进行数据分析。采用 Mann-Whitney U 检验比较组间差异。分析泪液参数之间的相关性。

结果

两组间年龄、logMAR 或眼压值无显著差异(p=0.480、0.077 和 0.146)。1 组和 2 组的平均 TBUT 分别为 5.1±2.9 和 9.5±4.2s(p<0.001)。1 组和 2 组的平均 TMH 分别为 237.9±108.9 和 324.3±158.9μm(p<0.001)。1 组的平均 TMA 明显低于 2 组(p<0.001;0.027±0.028 与 0.055±0.059mm²,分别)。Schirmer 值与 TMH 之间以及与 TMA 之间存在显著相关性(p=0.001 和<0.001)。

结论

Schirmer 值<5mm 的患者的 OCT 值明显较低。OCT 测量的泪膜半月板可可靠地用于诊断干眼症。