Canan Handan, Altan-Yaycioglu Rana, Ulas Burak, Sizmaz Selcuk, Coban-Karatas Muge
Department of Ophthalmology, Baskent University Faculty of Medicine, Adana Teaching & Medical Research Center , Adana , Turkey.
Curr Eye Res. 2014 Dec;39(12):1145-50. doi: 10.3109/02713683.2014.898311. Epub 2014 Apr 21.
New optical coherence tomography (OCT) devices with anterior segment module are able to measure the tear meniscus height (TMH) and tear meniscus area (TMA). Since the borders of the area and the height are to be marked by an examiner, the measurements are prone to be subjective. Herein, we aimed to evaluate the interexaminer reproducibility of the tear meniscus measurements with OCT.
In this prospective cross-sectional study, both eyes of 300 consecutive patients with and without dry eye were studied. Following routine ophthalmologic examination, tear-film OCT images were obtained. Two examiners, who were unaware of the other examiner's results, measured the TMH (i.e. the line connecting the intersection of the meniscus with the cornea/sclera and eyelid) and cross-sectional TMA. The reliability and correlation of the two examiners' results were assessed.
Four right and six left eyes were excluded from the final analysis, so a total of 590 eyes were evaluated. The mean difference of the two examiners' measurements of both eyes was -0.001 ± 0.027 mm(2) in TMA, and -21.29 ± 39.95 μm in TMH. An agreement between the two examiners was found regarding TMA and TMH measurements for right and left, as well as both eyes (Cronbach's alpha > 0.900, for all). Also, the correlation between both variables was high (inter item correlation matrix > 0.840, for all).
We showed a strong statistical agreement for both TMA and TMH measurements. According to our results, we believe that FD-OCT device is dependable in measuring the TMA and TMH values, given that its results are reproducible.
新型带有眼前节模块的光学相干断层扫描(OCT)设备能够测量泪膜半月板高度(TMH)和泪膜半月板面积(TMA)。由于区域和高度的边界需由检查者标记,测量容易受到主观因素影响。在此,我们旨在评估使用OCT测量泪膜半月板的检查者间重复性。
在这项前瞻性横断面研究中,对300例连续的干眼患者和非干眼患者的双眼进行了研究。在进行常规眼科检查后,获取泪膜OCT图像。两名彼此不知道对方结果的检查者测量了TMH(即连接半月板与角膜/巩膜和眼睑交点的线)和横断面TMA。评估了两名检查者结果的可靠性和相关性。
最终分析排除了4只右眼和6只左眼,因此共评估了590只眼。两名检查者对双眼的TMA测量平均差异为-0.001±0.027mm²,TMH测量平均差异为-21.29±39.95μm。发现两名检查者在右眼、左眼以及双眼的TMA和TMH测量方面存在一致性(所有Cronbach's alpha>0.900)。此外,两个变量之间的相关性很高(所有项间相关矩阵>0.840)。
我们显示出TMA和TMH测量在统计学上有很强的一致性。根据我们的结果,我们认为频域OCT设备在测量TMA和TMH值方面是可靠的,因为其结果具有可重复性。