J Refract Surg. 2013 May;29(5):356-8. doi: 10.3928/1081597X-20130129-03. Epub 2013 Mar 20.
To report the tomographic and biomechanical findings before and after treatment of a case of pressure-induced stromal keratopathy (PISK), which was misdiagnosed as diffuse lamellar keratitis (DLK).
A case report of a referred patient with supposed diagnosis of DLK after LASIK in the right eye. Scheimpflug-based corneal tomography and biomechanical assessment were provided by the Pentacam HR and CorVis ST (Oculus Optikgeräte GmbH, Wetzlar, Germany).
A layer of corneal opacity beneath the flap with a presumably fluid-filled interface area was observed on slit-lamp biomicroscopy. Scheimpflug image from Pentacam revealed a hyperreflective area underneath the flap interface. Goldmann applanation tonometry was 12 mm Hg, whereas CorVis intraocular pressure was 53.5 mm Hg with deformation amplitude of 0.42 mm. Two days after starting oral and topical ocular hypotensive therapy, CorVis intraocular pressure was 14 mm Hg and deformation amplitude was 1.02 mm.
Ocular hypertension in PISK was associated with lower deformation response, along with steepening and thickening of the cornea.
报告 1 例因 LASIK 术后压力诱导性基质角膜病变(PISK)而误诊为弥漫性层间角膜炎(DLK)的病例的治疗前后的断层和生物力学发现。
对右眼 LASIK 术后疑似 DLK 的转诊患者进行了病例报告。Pentacam HR 和 CorVis ST(Oculus Optikgeräte GmbH,德国韦茨拉尔)提供基于 Scheimpflug 的角膜断层和生物力学评估。
在裂隙灯生物显微镜下观察到瓣下有一层角膜混浊,可能是充满液体的界面区域。Pentacam 的 Scheimpflug 图像显示瓣下界面下有一个高反射区域。眼压计眼压为 12mmHg,而 CorVis 眼压为 53.5mmHg,变形幅度为 0.42mm。开始口服和局部降眼压治疗 2 天后,CorVis 眼压为 14mmHg,变形幅度为 1.02mm。
PISK 中的眼高压与角膜变陡和增厚相关的较低变形反应有关。