自动乳腺容积扫描仪(ABVS)解读的观察者间可靠性以及与手持乳腺超声(HHUS)、乳腺钼靶摄影和病理结果的一致性。

Interobserver reliability of automated breast volume scanner (ABVS) interpretation and agreement of ABVS findings with hand held breast ultrasound (HHUS), mammography and pathology results.

机构信息

University Breast Unit, Vossstrasse 9, 69115 Heidelberg, Germany.

出版信息

Eur J Radiol. 2013 Aug;82(8):e332-6. doi: 10.1016/j.ejrad.2013.03.005. Epub 2013 Mar 27.

Abstract

OBJECTIVES

Handheld breast ultrasound (HHUS) lacks standardization and reproducibility. The automated breast volume scanner (ABVS) could overcome this limitation. To analyze the interobserver reliability of ABVS and the agreement with HHUS, mammography and pathology is the aim of this study.

METHODS

All 42 study participants (=84 breasts) received an ABVS examination in addition to the conventional breast diagnostic work-up. 25 breasts (30%) showed at least one lesion. The scans were interpreted by six breast diagnostic specialists blinded to results of breast imaging and medical history. 32 lesions received histological work-up: 20 cancers were detected. We used kappa statistics to interpret agreement between examiners and diagnostic instruments.

RESULTS

On the basis of the Breast Imaging Reporting and Data System (BI-RADS) classification of the 84 breasts an agreement (defined as ≥4 of 6 examiners) was achieved in 63 cases (75%) (mk=0.35) and even improved when dichotomizing the interpretation in benign (BI-RADS 1, 2) and suspicious (BI-RADS 4, 5) to 98% (mk=0.52). Agreement of ABVS examination to HHUS, mammography and pathology was fair to substantial depending on the specific analysis.

CONCLUSIONS

The development of an ABVS seems to be a promising diagnostic method with a good interobserver reliability, as well as a comparable good test criteria as HHUS.

摘要

目的

手持式乳房超声(HHUS)缺乏标准化和可重复性。自动乳房容积扫描仪(ABVS)可以克服这一限制。本研究旨在分析 ABVS 的观察者间可靠性以及与 HHUS、乳房 X 光摄影和病理学的一致性。

方法

所有 42 名研究参与者(=84 只乳房)除了常规的乳房诊断程序外,还接受了 ABVS 检查。25 只乳房(30%)至少有一个病变。扫描由六名对乳房成像和病史结果均不知情的乳房诊断专家进行解读。32 个病变接受了组织学检查:发现 20 个癌症。我们使用 Kappa 统计来解释检查者和诊断仪器之间的一致性。

结果

根据 84 只乳房的乳腺影像报告和数据系统(BI-RADS)分类,在 63 例(75%)(mk=0.35)中达到了一致(定义为≥6 名检查者中的 4 名),当将解释分为良性(BI-RADS 1、2)和可疑(BI-RADS 4、5)时,一致性甚至提高到 98%(mk=0.52)。ABVS 检查与 HHUS、乳房 X 光摄影和病理学的一致性根据具体分析,从一般到很好。

结论

ABVS 的发展似乎是一种很有前途的诊断方法,具有良好的观察者间可靠性,以及与 HHUS 相当的良好测试标准。