Fuchs 内皮营养不良和后部板层角膜移植术后的眼压和角膜生物力学。

Intraocular pressure and corneal biomechanics in Fuchs' endothelial dystrophy and after posterior lamellar keratoplasty.

机构信息

Department of Ophthalmology, Aarhus University Hospital, Aarhus, Denmark.

出版信息

Acta Ophthalmol. 2014 Jun;92(4):350-4. doi: 10.1111/aos.12137. Epub 2013 Apr 23.

Abstract

PURPOSE

To evaluate the precision of techniques for measuring intraocular pressure (IOP) in corneas with presumably altered biomechanical properties.

METHODS

Intraocular pressure was measured with a Goldmann applanation tonometer (GAT), ocular response analyzer (ORA) and dynamic contour tonometer (DCT) in 70 eyes. Thirty-five eyes were normal corneas, 18 eyes had Fuchs' endothelial dystrophy, and 17 eyes had undergone Descemet's stripping automated endothelial keratoplasty (DSAEK) surgery. Corneal hysteresis (CH), corneal resistance factor (CRF) as well as central corneal thickness (CCT) were recorded with the ORA.

RESULTS

The measured cornea-corrected IOP using ORA was significantly higher than GAT in all three groups (p< 0.001). The DCT differed significantly from the Goldmann only in the Fuchs' group (p= 0.04). The Goldmann and DCT showed no significant between-group differences, whereas IOP measured with the ORA was different between groups (p< 0.001). CH in the DSEAK group differed significantly from the controls (p< 0.001), but there was no significant difference between the DSEAK and Fuchs' groups (p= 0.21). CCT did not differ significantly between the DSAEK and Fuchs' group (p= 0.47). However, both these groups differed significantly from the controls (p< 0.001).

CONCLUSION

Corneal hysteresis and CRF are reduced in Fuchs' endothelial dystrophy as well as after posterior lamellar keratoplasty. GAT and DCT seem to measure IOP correctly in patients with Fuchs' endothelial dystrophy as well as after posterior lamellar keratoplasty. Corneal-corrected IOP as measured with the ORA appears to overestimate IOP in patients with Fuchs' endothelial dystrophy as well as after posterior lamellar keratoplasty.

摘要

目的

评估在生物力学特性可能改变的角膜中测量眼压(IOP)的技术的精度。

方法

使用 Goldmann 压平眼压计(GAT)、眼反应分析仪(ORA)和动态轮廓眼压计(DCT)在 70 只眼中测量眼压。35 只眼为正常角膜,18 只眼为 Fuchs 内皮营养不良,17 只眼行 Descemet 氏膜撕开自动内皮角膜移植术(DSAEK)。用 ORA 记录角膜滞后(CH)、角膜阻力因子(CRF)和中央角膜厚度(CCT)。

结果

三组角膜校正后的 ORA 测量眼压均显著高于 GAT(p<0.001)。DCT 仅在 Fuchs 组与 Goldmann 存在显著差异(p=0.04)。Goldmann 和 DCT 之间在各组之间没有显著差异,而 ORA 测量的眼压在各组之间不同(p<0.001)。DSAEK 组的 CH 与对照组有显著差异(p<0.001),但与 Fuchs 组之间无显著差异(p=0.21)。CCT 在 DSAEK 组和 Fuchs 组之间无显著差异(p=0.47)。然而,这两组与对照组均有显著差异(p<0.001)。

结论

Fuchs 内皮营养不良和后板层角膜移植后,角膜滞后和 CRF 降低。GAT 和 DCT 似乎在 Fuchs 内皮营养不良和后板层角膜移植后能正确测量 IOP。ORA 测量的角膜校正后的 IOP 似乎高估了 Fuchs 内皮营养不良和后板层角膜移植后的 IOP。