The effect of increased corneal thickness after Descemet stripping endothelial keratoplasty (DSEK) on intraocular pressure (IOP) measurement has not been previously studied. It is uncertain if this increase in corneal thickness would artificially elevate IOP reading by Goldmann tonometry [GAT] (Haag-Streit, Konig, Switzerland). Therefore the effect of DSEK-related thick cornea on IOP measurement was investigated using three different techniques.
Prospective cross-sectional study.
Participants were recruited from a single tertiary referral center. Fifty eyes of 38 patients with successful DSEK at least three months prior to testing were evaluated. At the time of the study, none of the participants had clinically detectable corneal edema. IOP was measured with GAT, pneumatonometry, and dynamic contour tonometry (DCT) in an unmasked randomized sequence. Central corneal thickness (CCT) was measured by ultrasonic pachymetry.
Mean CCT was 701 +/- 68 microm. The mean IOP +/- standard deviation (SD) was 15.9 +/- 4.9 mm Hg for GAT, 20.3 +/- 4.5 mm Hg for pneumatonometry, and 19.8 +/- 4.4 mm Hg for DCT. Pneumatonometry and DCT IOP measurements were significant higher than GAT (P < .01). In contrast, the difference between pneumatonometry and DCT readings was not statistically significant (P = .28). The correlations between IOP and corneal thickness were not significant in this cohort (P > .05).
Falsely elevated GAT, as expected in thick corneas, was not demonstrated after DSEK. High IOP reading by GAT therefore should raise suspicion of elevated IOP in DSEK eyes.
以往尚未研究过角膜后弹力层剥除内皮角膜移植术(DSEK)后角膜厚度增加对眼压(IOP)测量的影响。目前尚不确定这种角膜厚度的增加是否会通过戈德曼眼压计(GAT, Haag-Streit,Konig,瑞士)人为地提高眼压读数。因此,使用三种不同技术研究了DSEK相关的角膜增厚对眼压测量的影响。
前瞻性横断面研究。
研究对象来自单一的三级转诊中心。对38例在测试前至少三个月成功进行DSEK的患者的50只眼进行评估。在研究时,所有参与者均无临床可检测到的角膜水肿。以未遮盖的随机顺序用GAT、气压眼压计和动态轮廓眼压计(DCT)测量眼压。通过超声测厚法测量中央角膜厚度(CCT)。
平均CCT为701±68微米。GAT测量的平均IOP±标准差(SD)为15.9±4.9 mmHg,气压眼压计测量为20.3±4.5 mmHg,DCT测量为19.8±4.4 mmHg。气压眼压计和DCT测量的眼压显著高于GAT(P <.01)。相比之下,气压眼压计和DCT读数之间的差异无统计学意义(P = 0.28)。在该队列中,眼压与角膜厚度之间的相关性不显著(P>.05)。
DSEK后未出现厚角膜预期的GAT假性升高。因此,GAT测得的高眼压读数应引起对DSEK眼眼压升高的怀疑。