We assessed the analytical performance of the TSH and FT4 assays on ADVIA Centaur in a multicenter national evaluation.
A precision study and a method comparison were performed. Reference values stated by the manufacturer were checked from 379 normal subjects.
For TSH and FT4, the intra-assay CVs were below 2.3 and 5.2%, respectively, and the inter-assay CVs below 4.4% and 7.2%, respectively. Therefore, the precision and reproducibility were acceptable. Bland-Altman bias plots revealed good correlation and agreement with Cobas assays. TSH and FT4 data yielded reference ranges of 0.64-3.24 mIU/L and 10.5-18.9 pmol/L, respectively.
These assays demonstrate reliable characteristics. The reference ranges obtained can be used for interpretation of thyroid function.
我们评估了 ADVIA Centaur 上 TSH 和 FT4 检测的分析性能,这是一项多中心的全国性评估。
进行了精密度研究和方法比较。检查了 379 名正常受试者的制造商规定的参考值。
对于 TSH 和 FT4,其日内 CV 分别低于 2.3%和 5.2%,而日间 CV 分别低于 4.4%和 7.2%。因此,精密度和重现性是可以接受的。Bland-Altman 偏倚图显示与 Cobas 检测具有良好的相关性和一致性。TSH 和 FT4 的参考范围分别为 0.64-3.24 mIU/L 和 10.5-18.9 pmol/L。
这些检测方法具有可靠的特征。获得的参考范围可用于解释甲状腺功能。