Adv Orthop. 2013;2013:710621. doi: 10.1155/2013/710621. Epub 2013 Apr 11.
Multiple factors have been identified as contributing to polyethylene wear and debris generation of the acetabular lining. Polyethylene wear is the primary limiting factor in the functional behavior and consequent longevity of a total hip arthroplasty (THA). This retrospective study reviewed the clinical and radiographic data of 77 consecutive THAs comparing in vivo polyethylene wear of two similar acetabular cup liners. Minimum follow-up was 7 years (range 7-15). The incidence of measurable wear in a group of machined liners sterilized with ethylene oxide and composed of GUR 1050 stock resin was significantly higher (61%) than the compression-molded, GUR 1020, O2-free gamma irradiation sterilized group (24%) (P = 0.0004). Clinically, at a 9-year average followup, both groups had comparable HHS scores and incidence of thigh or groin pain, though the machined group had an increased incidence of osteolysis and annual linear wear rate.
多种因素已被确定为导致髋臼衬里聚乙烯磨损和碎屑产生的原因。聚乙烯磨损是全髋关节置换术(THA)功能行为及后续使用寿命的主要限制因素。这项回顾性研究回顾了77例连续THA的临床和影像学数据,比较了两种相似髋臼杯衬里的体内聚乙烯磨损情况。最短随访时间为7年(范围7 - 15年)。一组用环氧乙烷灭菌、由GUR 1050原料树脂制成的机加工衬里的可测量磨损发生率(61%)显著高于压缩成型、GUR 1020、无氧伽马射线灭菌组(24%)(P = 0.0004)。临床上,在平均9年的随访中,两组的髋关节 Harris 评分(HHS)及大腿或腹股沟疼痛发生率相当,不过机加工组的骨溶解发生率和年线性磨损率有所增加。