Coronary CT angiography using the second-generation 320-detector row CT: assessment of image quality and radiation dose in various heart rates compared with the first-generation scanner.

机构信息

Department of Radiology, The University of Tokyo, 7-3-1 Hongo, Bunkyo-ku, Tokyo, 113-0033, Japan,

出版信息

Int J Cardiovasc Imaging. 2013 Oct;29(7):1613-8. doi: 10.1007/s10554-013-0238-2. Epub 2013 May 17.

Abstract

To assess the image quality and radiation dose reduction in various heart rates in coronary CT angiography using the second-generation 320-detector row CT compared with the first-generation CT. Ninety-six patients were retrospectively included. The first 48 patients underwent coronary CT angiography with the first-generation 320-detector row CT, while the last 48 patients underwent with the second-generation CT. Subjective image quality was graded using a 4-point scale (4, excellent; 1, unable to evaluate). Image noise and contrast-to-noise ratio were also analyzed. Subgroup analysis was performed based on the heart rate. The mean effective dose was derived from the dose length product multiplied by a conversion coefficient for the chest (κ = 0.014 mSv × mGy(-1) × cm(-1)). The overall subjective image quality score showed no significant difference (3.66 vs 3.69, respectively, p = 0.25). The image quality score of the second-generation group tended to be higher than that of the first-generation group in the 66- to 75-bpm subgroup (3.36 vs 3.53, respectively, p = 0.07). No significant difference was observed in image noise and contrast-to-noise ratio. The overall radiation dose reduced by 24 % (3.3 vs 2.5 mSv, respectively, p = 0.03), and the reduction was substantial in patients with higher heart rate (66- to 75-bpm, 4.3 vs 2.2 mSv, respectively, p = 0.009; >75 bpm, 8.2 vs 3.7 mSv, respectively, p = 0.005). The second-generation 320-detector row CT could maintain the image quality while reducing the radiation dose in coronary CT angiography. The dose reduction was larger in patients with higher heart rate.

摘要

评估第二代 320 排 CT 与第一代 CT 相比在不同心率下冠状动脉 CT 血管造影的图像质量和辐射剂量降低。回顾性纳入 96 例患者。前 48 例患者行第一代 320 排 CT 冠状动脉 CT 血管造影,后 48 例患者行第二代 CT。采用 4 分制(4 分,极好;1 分,无法评估)对主观图像质量进行分级。还分析了图像噪声和对比噪声比。根据心率进行亚组分析。有效剂量由剂量长度乘积乘以胸部转换系数(κ=0.014 mSv×mGy(-1)×cm(-1))得出。总体主观图像质量评分无显著差异(分别为 3.66 和 3.69,p=0.25)。在 66-75 bpm 亚组中,第二代组的图像质量评分倾向于高于第一代组(分别为 3.36 和 3.53,p=0.07)。图像噪声和对比噪声比无显著差异。总体辐射剂量降低 24%(分别为 3.3 和 2.5 mSv,p=0.03),在心率较高的患者中降低幅度较大(66-75 bpm,分别为 4.3 和 2.2 mSv,p=0.009;>75 bpm,分别为 8.2 和 3.7 mSv,p=0.005)。第二代 320 排 CT 可在保持冠状动脉 CT 血管造影图像质量的同时降低辐射剂量。心率较高的患者剂量降低幅度更大。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验