Ferris B D
Westminster Hospital, London, UK.
J Exp Pathol (Oxford). 1990 Jun;71(3):367-73.
An attempt has been made to refine the quantitative evaluation of the tissue response to a prosthesis using the goat hip as an experimental model. Total joint replacement was performed in six goats; in three the cup was made of conventional ultrahigh molecular weight polyethylene and in three a new silane cross-linked ultrahigh molecular weight polyethylene was used. The pseudocapsule of the hips was examined histologically at 15, 19 and 24 months following surgery. Small and large polyethylene debris were visualized using semi-crossed polars and were counted. Giant cells were counted after staining with toluidine blue as were macrophages which were identified by dense staining after the acid phosphatase reaction. The amount of small birefringent debris was closely related to the number of macrophages in the pseudocapsule. This method of assessment may be helpful in comparing the tissue reaction of different polymers.
已尝试以山羊髋关节作为实验模型,对组织对假体的反应进行定量评估。对六只山羊进行了全关节置换;其中三只使用传统超高分子量聚乙烯制成的髋臼杯,另外三只使用新型硅烷交联超高分子量聚乙烯。在术后15、19和24个月对髋关节的假包膜进行组织学检查。使用半交叉偏振光观察并计数大小不同的聚乙烯碎片。用甲苯胺蓝染色后计数巨细胞,酸磷酸酶反应后通过密集染色鉴定巨噬细胞。假包膜中双折射小碎片的数量与巨噬细胞的数量密切相关。这种评估方法可能有助于比较不同聚合物的组织反应。