Department of Chemistry, University of Michigan, Ann Arbor, MI 48109, USA.
Biosens Bioelectron. 2013 Nov 15;49:204-9. doi: 10.1016/j.bios.2013.05.014. Epub 2013 May 16.
Miniature enzyme-based amperometric and coulometric glucose sensors were fabricated and applied to measure tear glucose concentrations in anesthetized rabbits. Without perturbing the eyeball, 3 µL of tear fluid was sampled from the marginal conjunctiva under the lower eyelid of anesthetized rabbits at various time points via a microliter glass capillary tube, and the miniature sensors were then inserted into the volume of collected tear fluids within the capillaries for detection. Intravenous bolus doses of insulin were administrated to the rabbits to lower the elevated blood glucose concentrations caused by anesthesia over the 7h test periods. A significant correlation was found between tear and blood glucose levels for multiple rabbits, suggesting that electrochemical sensor-based tear glucose measurements may be a potential supplementary method for point-of-care glucose monitoring.
微型酶基安培和库仑葡萄糖传感器被制备并应用于测量麻醉兔的泪液葡萄糖浓度。在不干扰眼球的情况下,通过微量玻璃毛细管从麻醉兔下眼睑的边缘结膜处采集 3µL 的泪液样品,然后将微型传感器插入毛细管内采集的泪液体积中进行检测。通过静脉推注胰岛素的方式给兔子给药,以降低麻醉引起的高血糖浓度,在 7 小时的测试期间内。对多个兔子进行了研究,发现泪液和血糖水平之间存在显著相关性,这表明基于电化学传感器的泪液葡萄糖测量可能是一种潜在的即时血糖监测补充方法。