Department of Chemistry, The University of Michigan, Ann Arbor, Michigan 48109-1055, United States.
Anal Chem. 2011 Nov 1;83(21):8341-6. doi: 10.1021/ac201700c. Epub 2011 Oct 12.
An amperometric needle-type electrochemical glucose sensor intended for tear glucose measurements is described and employed in conjunction with a 0.84 mm i.d. capillary tube to collect microliter volumes of tear fluid. The sensor is based on immobilizing glucose oxidase on a 0.25 mm o.d. platinum/iridium (Pt/Ir) wire and anodically detecting the liberated hydrogen peroxide from the enzymatic reaction. Inner layers of Nafion and an electropolymerized film of 1,3-diaminobenzene/resorcinol greatly enhance the selectivity for glucose over potential interferences in tear fluid, including ascorbic acid and uric acid. Further, the new sensor is optimized to achieve very low detection limits of 1.5 ± 0.4 μM of glucose (S/N = 3) that is required to monitor glucose levels in tear fluid with a glucose sensitivity of 0.032 ± 0.02 nA/μM (n = 6). Only 4-5 μL of tear fluid in the capillary tube is required when the needle sensor is inserted into the capillary. The glucose sensor was employed to measure tear glucose levels in anesthetized rabbits over an 8 h period while also measuring the blood glucose values. A strong correlation between tear and blood glucose levels was found, suggesting that measurement of tear glucose is a potential noninvasive substitute for blood glucose measurements, and the new sensor configuration could aid in conducting further research in this direction.
一种用于泪液葡萄糖测量的安培型针式电化学葡萄糖传感器被描述并与 0.84mm 内径的毛细管一起使用,以收集微升体积的泪液。该传感器基于将葡萄糖氧化酶固定在 0.25mm 外径的铂/铱(Pt/Ir)丝上,并通过阳极检测酶反应中释放的过氧化氢。Nafion 的内层和 1,3-二氨基苯/间苯二酚的电聚合膜极大地提高了传感器对泪液中潜在干扰物质(包括抗坏血酸和尿酸)的选择性。此外,新传感器经过优化,可实现非常低的检测限,对于监测泪液中的葡萄糖水平,检测限为 1.5±0.4μM(S/N=3),葡萄糖灵敏度为 0.032±0.02nA/μM(n=6)。当将针式传感器插入毛细管时,仅需要毛细管中的 4-5μL 泪液。该葡萄糖传感器被用于在麻醉兔子中测量 8 小时的泪液葡萄糖水平,同时测量血糖值。发现泪液和血糖水平之间存在很强的相关性,表明测量泪液葡萄糖是血糖测量的一种潜在非侵入性替代方法,并且新的传感器配置可以有助于进一步进行这方面的研究。