Suppr超能文献

多焦点人工晶状体的光学性能。

Optical performance of multifocal intraocular lenses.

作者信息

Holladay J T, Van Dijk H, Lang A, Portney V, Willis T R, Sun R, Oksman H C

机构信息

University of Texas Medical School, Department of Ophthalmology, Hermann Eye Center, Houston 77030.

出版信息

J Cataract Refract Surg. 1990 Jul;16(4):413-22. doi: 10.1016/s0886-3350(13)80793-6.

Abstract

The optical performance of one monofocal and five multifocal lenses was evaluated in the laboratory and photographically. The laboratory testing included determination of the modulation transfer function (MTF), through focus response (TFR), resolution efficiency, and Strehl ratio of each lens. The photographic testing included photographs of the Regan high contrast acuity chart at ten feet with clearest focus and 18 additional photographs in which the image was defocused using minus trial lenses in 0.25 diopter increments. A color photograph of the Kodak color chart was also taken using each lens. All testing was conducted using a 3 mm artificial pupil under ideal implant conditions with no decentration or tilt. The laboratory and photographic results demonstrate that all the multifocal lenses had a two- to three-fold increase in the depth of field with at least a 50% lower contrast in the retinal image. The photographic testing revealed a one to two line better resolution limit with the monofocal lens, which corresponded to the 12% to 41% better MTF cut-off value with the monofocal lens by laboratory testing. The measured resolution efficiencies of all six lenses were comparable. The color photographs revealed color mixing of adjacent colors with the multifocal lenses, whereas the colors appeared unchanged from the original with the monofocal lens.

摘要

在实验室中并通过摄影对一片单焦点镜片和五片多焦点镜片的光学性能进行了评估。实验室测试包括测定每片镜片的调制传递函数(MTF)、全焦距响应(TFR)、分辨率效率和斯特列尔比。摄影测试包括在十英尺距离拍摄里根高对比度视力表,使焦点最清晰,以及另外18张照片,其中使用负球镜试片以0.25屈光度的增量使图像散焦。还使用每片镜片拍摄了柯达彩色图表的彩色照片。所有测试均在理想的植入条件下,使用3毫米人工瞳孔进行,无偏心或倾斜。实验室和摄影结果表明,所有多焦点镜片的景深增加了两到三倍,而视网膜图像中的对比度至少降低了50%。摄影测试显示,单焦点镜片的分辨率极限要好一到两行,这与实验室测试中该单焦点镜片的MTF截止值高12%至41%相对应。所测量的所有六片镜片的分辨率效率相当。彩色照片显示,多焦点镜片存在相邻颜色的混色现象,而单焦点镜片的颜色与原始颜色相比没有变化。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验