Clinical evaluation of a new continuous intraarterial blood gas monitoring system in the intensive care setting.

机构信息

Department of Intensive and Critical Care Medicine, Jichi Medical School, 3311-1 Yakushiji, Minamikawachi, Kawachi-gun, 329-04, Tochigi, Japan.

出版信息

J Anesth. 1996 Sep;10(3):163-9. doi: 10.1007/BF02471384.

Abstract

The present study was designed to evaluate a new continuous intraarterial blood gas monitoring system under routine clinical intensive care conditions. Nine mechanically ventilated adult patients were enrolled in this study. A multiparameter intravascular sensor was inserted into the radial or dorsalis pedis artery through a 20-gauge cannula in each patient. The accuracy of the sensor for pH, Pco2, and Po2 values was evaluated by comparing the data simultaneously obtained from the monitoring system and from conventional blood gas analysis. Measurements were performed for 3 days for each sensor. A total of 62 blood samples were obtained for comparison. The ranges of measured variables were: pH 7.185-7.602, Pco2, 28.8-68.5 mmHg, and Po2 45.2-542.4 mmHg. The overall bias ±precision values were 0.002±0.018 for pH units, 0.53±2.04mmHg for Pco2, and -1.62±20.00 mmHg for Po2. In clinically important ranges of Po2, less than 200 mmHg in particular, the bias and precision values were -2.25±6.48 mmHg in the range of less than 100mmHg, and 0.98±14.38 mmHg in the range of 100-200 mmHg. Variations of sensor accuracy as a function of elapsed time were within the clinically acceptable range throughout the study period. These findings suggest that this new device is sufficiently useful for routine clinical settings.

摘要

本研究旨在评估一种新的连续动脉血气监测系统在常规临床重症监护条件下的应用。本研究共纳入 9 例机械通气的成年患者。每位患者经 20G 套管针将多参数血管内传感器插入桡动脉或足背动脉。通过同时比较监测系统和常规血气分析获得的数据,评估传感器对 pH、Pco2 和 Po2 值的准确性。每个传感器的测量时间为 3 天,共获得 62 份血样进行比较。测量变量的范围为:pH 值 7.185-7.602,Pco2 28.8-68.5mmHg,Po2 45.2-542.4mmHg。总体偏差±精度值分别为 pH 单位 0.002±0.018,Pco2 0.53±2.04mmHg,Po2-1.62±20.00mmHg。在 Po2 的临床重要范围内,特别是在 200mmHg 以下,100mmHg 以下的偏差和精度值分别为-2.25±6.48mmHg,100-200mmHg 的偏差和精度值分别为 0.98±14.38mmHg。整个研究期间,传感器精度随时间变化的偏差均在临床可接受范围内。这些发现表明,该新型设备在常规临床环境中具有足够的应用价值。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验