应用谱域光学相干断层扫描比较手动与自动的椎间盘边缘勾画。

The comparison of manual vs automated disc margin delineation using spectral-domain optical coherence tomography.

机构信息

Bascom Palmer Eye Institute, Department of Ophthalmology, University of Miami Miller School of Medicine, Palm Beach Gardens, FL, USA.

出版信息

Eye (Lond). 2013 Oct;27(10):1180-7. doi: 10.1038/eye.2013.163. Epub 2013 Aug 2.

Abstract

AIMS

To examine the impact of manual vs automated disc margin delineation on optic nerve head (ONH) and retinal nerve fiber layer (RNFL) parameters using spectral-domain optical coherence tomography (SDOCT).

METHODS

A prospective cohort study consisting of normal, glaucoma suspect (GS) and glaucoma patients who underwent ONH and RNFL measurements using SDOCT technology (RTVue; Optovue Inc.). The retinal pigment epithelium/Bruch's membrane (RPE/BM) complex end points were automatically determined first, and were manually redefined subsequently. Analysis of variance, coefficient of variation (COV), concordance correlation coefficient (CCC), and Bland-Altman plots were used for the analyses.

RESULTS

Ninety-nine eyes of 50 subjects (age 68±10 years) consisting of 36 glaucoma, 56 GS, and 7 normal eyes were included. The RNFL thickness measurements were similar (P>0.05) between the two methods of demarcation, except for the inferior-nasal sector (P=0.04). For the ONH measurements, the cup-to-disc (C/D) ratio and rim area showed significant differences between the two methods (P<0.001). COV/CCC values for the ONH parameters were as follows: cup area 17.6%/0.88; cup volume 7.4%/0.91; average C/D ratio 18.1%/0.78; rim area 25.3%/0.69; and rim volume 42.6%/0.71, respectively. CCC/COV values for the RNFL parameters were as follows: average 2.1%/0.98; inferior-temporal quadrant 8.1%/0.79; inferior-nasal quadrant INQ quadrant 12.6%/0.67; SNQ quadrant 7.8%/0.83; and STQ quadrant 7.8%/0.88, respectively.

CONCLUSION

An overall high agreement and moderate-substantial concordance was observed between the demarcation methods. Automated disc margin delineation of SDOCT can be used reliably in clinical practice.

摘要

目的

使用频域光学相干断层扫描(SDOCT)检查手动与自动视盘边缘描绘对视神经头(ONH)和视网膜神经纤维层(RNFL)参数的影响。

方法

前瞻性队列研究包括接受 SDOCT 技术(RTVue; Optovue Inc.)进行 ONH 和 RNFL 测量的正常、青光眼疑似(GS)和青光眼患者。首先自动确定视网膜色素上皮/布鲁赫膜(RPE/BM)复合终点,然后手动重新定义。使用方差分析、变异系数(COV)、一致性相关系数(CCC)和 Bland-Altman 图进行分析。

结果

纳入 50 名受试者(年龄 68±10 岁)的 99 只眼,包括 36 只青光眼、56 只 GS 和 7 只正常眼。两种描绘方法的 RNFL 厚度测量值相似(P>0.05),除了下鼻象限(P=0.04)。对于 ONH 测量,两种方法之间杯盘比(C/D)和边缘面积有显著差异(P<0.001)。ONH 参数的 COV/CCC 值如下:杯面积 17.6%/0.88;杯容积 7.4%/0.91;平均 C/D 比 18.1%/0.78;边缘面积 25.3%/0.69;边缘容积 42.6%/0.71。RNFL 参数的 CCC/COV 值如下:平均 2.1%/0.98;下颞象限 8.1%/0.79;下鼻象限 INQ 象限 12.6%/0.67;SNQ 象限 7.8%/0.83;和 STQ 象限 7.8%/0.88。

结论

两种描绘方法之间观察到总体高度一致和中度实质性一致。SDOCT 的自动视盘边缘描绘可在临床实践中可靠使用。

相似文献

3
Application of optical coherence tomography in glaucoma suspect eyes.光学相干断层扫描在青光眼可疑眼中的应用。
Clin Exp Optom. 2012 Jan;95(1):78-88. doi: 10.1111/j.1444-0938.2011.00654.x. Epub 2011 Oct 10.

引用本文的文献

本文引用的文献

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验