Department of Radiology, The University of Texas Medical Branch, Galveston, TX 77550, USA.
Med Phys. 2013 Aug;40(8):082302. doi: 10.1118/1.4816306.
To investigate the utility of five different standard measurement methods for determining image uniformity for partially parallel imaging (PPI) acquisitions in terms of consistency across a variety of pulse sequences and reconstruction strategies.
Images were produced with a phantom using a 12-channel head matrix coil in a 3T MRI system (TIM TRIO, Siemens Medical Solutions, Erlangen, Germany). Images produced using echo-planar, fast spin echo, gradient echo, and balanced steady state free precession pulse sequences were evaluated. Two different PPI reconstruction methods were investigated, generalized autocalibrating partially parallel acquisition algorithm (GRAPPA) and modified sensitivity-encoding (mSENSE) with acceleration factors (R) of 2, 3, and 4. Additionally images were acquired with conventional, two-dimensional Fourier imaging methods (R=1). Five measurement methods of uniformity, recommended by the American College of Radiology (ACR) and the National Electrical Manufacturers Association (NEMA) were considered. The methods investigated were (1) an ACR method and a (2) NEMA method for calculating the peak deviation nonuniformity, (3) a modification of a NEMA method used to produce a gray scale uniformity map, (4) determining the normalized absolute average deviation uniformity, and (5) a NEMA method that focused on 17 areas of the image to measure uniformity. Changes in uniformity as a function of reconstruction method at the same R-value were also investigated. Two-way analysis of variance (ANOVA) was used to determine whether R-value or reconstruction method had a greater influence on signal intensity uniformity measurements for partially parallel MRI.
Two of the methods studied had consistently negative slopes when signal intensity uniformity was plotted against R-value. The results obtained comparing mSENSE against GRAPPA found no consistent difference between GRAPPA and mSENSE with regard to signal intensity uniformity. The results of the two-way ANOVA analysis suggest that R-value and pulse sequence type produce the largest influences on uniformity and PPI reconstruction method had relatively little effect.
Two of the methods of measuring signal intensity uniformity, described by the (NEMA) MRI standards, consistently indicated a decrease in uniformity with an increase in R-value. Other methods investigated did not demonstrate consistent results for evaluating signal uniformity in MR images obtained by partially parallel methods. However, because the spatial distribution of noise affects uniformity, it is recommended that additional uniformity quality metrics be investigated for partially parallel MR images.
研究在不同脉冲序列和重建策略下,五种不同的标准测量方法在部分并行成像(PPI)采集的图像均匀性方面的一致性,以确定其在部分并行成像中的实用性。
在 3T MRI 系统(西门子医疗解决方案公司的 TIM TRIO)中,使用 12 通道头部矩阵线圈在体模上生成图像。评估了使用回波平面、快速自旋回波、梯度回波和平衡稳态自由进动脉冲序列生成的图像。研究了两种不同的 PPI 重建方法,即广义自校准部分并行采集算法(GRAPPA)和改进的灵敏度编码(mSENSE),加速因子(R)分别为 2、3 和 4。此外,还使用传统的二维傅里叶成像方法(R=1)采集图像。考虑了美国放射学院(ACR)和美国电器制造商协会(NEMA)推荐的五种均匀性测量方法。研究的方法包括:(1)一种用于计算峰值偏差非均匀性的 ACR 方法和(2)一种 NEMA 方法;(3)一种用于生成灰度均匀性图的 NEMA 方法的修改版;(4)确定归一化绝对平均偏差均匀性;(5)一种关注图像 17 个区域以测量均匀性的 NEMA 方法。还研究了在相同 R 值下,重建方法变化对均匀性的影响。采用双因素方差分析(ANOVA)来确定 R 值或重建方法对部分并行 MRI 信号强度均匀性测量的影响更大。
当信号强度均匀性与 R 值作图时,研究的两种方法的斜率均为负。对 mSENSE 与 GRAPPA 进行比较的结果发现,在信号强度均匀性方面,GRAPPA 与 mSENSE 之间没有一致的差异。双因素方差分析的结果表明,R 值和脉冲序列类型对均匀性的影响最大,而 PPI 重建方法的影响相对较小。
两种描述(NEMA)MRI 标准的信号强度均匀性测量方法一致表明,随着 R 值的增加,均匀性降低。其他研究的方法在评估部分并行方法获得的磁共振图像的信号均匀性时没有显示出一致的结果。然而,由于噪声的空间分布会影响均匀性,因此建议对部分并行磁共振图像的均匀性质量指标进行进一步研究。