Assessment of the axial rotation of the pelvis with the EOS® imaging system: intra- and inter-observer reproducibility and accuracy study.

作者信息

Rousseau Marc-Antoine, Brusson Adrien, Lazennec Jean-Yves

机构信息

Department of Orthopaedic Surgery, Hôpital Avicenne-Université Paris 13, Bobigny, France,

出版信息

Eur J Orthop Surg Traumatol. 2014 Aug;24(6):891-5. doi: 10.1007/s00590-013-1281-3. Epub 2013 Aug 13.

Abstract

PURPOSE

A source of radiological bias occurs when the axial rotation of the pelvis is disregarded in hip and spine biomechanics analyses. The EOS imaging system (EOS Imaging, France) offers the possibility of detecting and measuring the axial rotation of bones. Reproducibility and accuracy have not been documented in the case of the pelvis.

METHODS

A dry pelvis has been X-rayed with the EOS system every 5° from 30° left to 30° right according to a laser line reference goniometer. Three observers have measured the rotation. One observer did it 3 times. The intra- and inter-observer reproducibility and the accuracy have been calculated using the root mean square standard deviation calculation. The relationship between the axial rotation and the offset between the left and right acetabulae on the lateral view was investigated.

RESULTS

The 95 % CI was ±0.23° for the intra-observer and ±0.33° for the inter-observer reliability. The mean of signed differences between the software calculation and the actual axial rotation of the pelvis was -0.39° (SD 0.77°). The lateral acetabular offset was proportional to the sin of the rotation. Approximately, 30 mm offset corresponded to about 10° rotation.

CONCLUSIONS

The 3D slot scanning imaging system demonstrated significant reproducibility and accuracy for the assessment of the axial rotation of the pelvis.