Suppr超能文献

再生与未再生氧化纤维素止血剂的比较。

Comparison of regenerated and non-regenerated oxidized cellulose hemostatic agents.

作者信息

Lewis K M, Spazierer D, Urban M D, Lin L, Redl H, Goppelt A

机构信息

Baxter Healthcare Corporation, One Baxter Parkway, Deerfield, 60015 IL USA.

出版信息

Eur Surg. 2013;45(4):213-220. doi: 10.1007/s10353-013-0222-z. Epub 2013 Jul 4.

Abstract

BACKGROUND

Oxidized cellulose is a well known and widely used surgical hemostat. It is available in many forms, but manufactured using either a nonregenerated or regenerated process.

OBJECTIVE

This study compares the fiber structure, pH in solution, bactericidal effectiveness, and hemostatic effectiveness of an oxidized nonregenerated cellulose (ONRC; Traumastem) and an oxidized regenerated cellulose (ORC; Surgicel Original).

METHODS

In vitro, fiber structures were compared using scanning electron microscopy, pH of phosphate buffer solution (PBS) and human plasma were measured after each cellulose was submerged, and bactericidal effect was measured by plating each cellulose with four bacteria. In vivo, time to hemostasis and hemostatic success were compared using a general surgery nonheparinized porcine liver abrasion model and a peripheral vascular surgery heparinized leporine femoral vessel bleeding model.

RESULTS

Ultrastructure of ONRC fiber is frayed, while ORC is smooth. ORC pH is statistically more acidic than ONRC in PBS, but equal in plasma. No difference in bactericidal effectiveness was observed. In vivo, ONRC provided superior time to hemostasis relative to ORC (211.2 vs 384.6 s,  = 60/group) in the general surgery model; and superior hemostatic success relative to ORC at 30 (60 vs. 15 %; OR: 13.5; 95 % CI: 3.72-49.1,  = 40/group), 60 (85 vs. 37.5 %; OR: 12.3; 95 % CI: 3.66-41.6), and 90 s (97.5 vs 70.0 %; OR: 21.1, 95 % CI: 2.28-195.9) in the peripheral vascular model.

CONCLUSION

ONRC provides superior hemostasis and equivalent bactericidal effectiveness relative to ORC, which is likely due to its fiber structure than acidity.

摘要

背景

氧化纤维素是一种广为人知且广泛应用的外科止血剂。它有多种形式,但采用非再生或再生工艺制造。

目的

本研究比较了氧化非再生纤维素(ONRC;Traumastem)和氧化再生纤维素(ORC;Surgicel Original)的纤维结构、溶液中的pH值、杀菌效果和止血效果。

方法

在体外,使用扫描电子显微镜比较纤维结构,将每种纤维素浸入后测量磷酸盐缓冲溶液(PBS)和人血浆的pH值,通过将每种纤维素与四种细菌接种来测量杀菌效果。在体内,使用普通外科非肝素化猪肝擦伤模型和外周血管外科肝素化兔股血管出血模型比较止血时间和止血成功率。

结果

ONRC纤维的超微结构有磨损,而ORC则很光滑。在PBS中,ORC的pH值在统计学上比ONRC更酸性,但在血浆中相等。未观察到杀菌效果的差异。在体内,在普通外科模型中,ONRC相对于ORC提供了更短的止血时间(211.2对384.6秒,每组=60);在外周血管模型中,在30秒(60%对15%;OR:13.5;95%CI:3.72-49.1,每组=40)、60秒(85%对37.5%;OR:12.3;95%CI:3.66-41.6)和90秒(97.5%对70.0%;OR:21.1,95%CI:2.28-195.9)时,ONRC相对于ORC有更高的止血成功率。

结论

相对于ORC,ONRC具有更好的止血效果和相当的杀菌效果,这可能是由于其纤维结构而非酸度。