AJR Am J Roentgenol. 2013 Sep;201(3):W511-5. doi: 10.2214/AJR.12.9571.
Many indications for breast MRI exist. MRI screening can identify preinvasive breast cancer in women at high risk and in that regard is superior to mammography and ultrasound. Quality control standards exist for mammography and breast ultra-sound screening with phantoms designed specifically for this purpose. Given the growing importance of breast MRI, we propose the development of a breast MRI phantom for quality control purposes.
A breast phantom with dual cavities containing water and fat was developed. A resolution plate inside the phantom contains various shapes ranging in size from 1 to 20 mm. Twenty studies of the phantom were performed with a 1.5-T system. STIR, T1-weighted fat-suppressed, and T2-weighted sequences were completed. Relaxation times of water and fat, number of step shapes resolved on STIR and T2-weighted images, number of circles resolved on T2-weighted images, and the diameter of a 20-mm circle on T1-weighted fat-suppressed images were recorded.
On STIR images the TR of fat was 238.70±96.31 ms and of water was 1231.92±399.14 ms. On T2-weighted images the TR of fat was 778.73±62.60 ms and of water was 1737.60±121.63 ms. On STIR images, steps 3 mm and larger were visualized in 95% of instances. On T2-weighted images steps 3 mm and larger were seen in all instances. Measurements of a 20-mm circle were 19±0.3 mm.
The proposed breast MRI phantom can be used to obtain reproducible measurements and allows implementation of quality control measures for a modality that is being increasingly used.
乳腺 MRI 有多种适应证。与乳腺钼靶和超声相比,MRI 筛查可在高危人群中发现早期浸润性乳腺癌,因此具有优势。乳腺钼靶和超声筛查存在质量控制标准,且有专门为此目的设计的成像体模。鉴于乳腺 MRI 的重要性日益增加,我们建议开发一种用于质量控制的乳腺 MRI 体模。
设计了一种具有双腔的乳腺体模,其中包含水和脂肪。体模内的分辨率板包含各种尺寸从 1 到 20 毫米的形状。在 1.5T 系统上对该体模进行了 20 次研究。完成了 STIR、T1 加权脂肪抑制和 T2 加权序列。记录了水和脂肪的弛豫时间、STIR 和 T2 加权图像上分辨的阶梯形状数量、T2 加权图像上分辨的圆形数量以及 T1 加权脂肪抑制图像上 20 毫米圆形的直径。
在 STIR 图像上,脂肪的 TR 为 238.70±96.31ms,水的 TR 为 1231.92±399.14ms。在 T2 加权图像上,脂肪的 TR 为 778.73±62.60ms,水的 TR 为 1737.60±121.63ms。在 STIR 图像上,95%的情况下可以观察到 3mm 及以上的台阶。在 T2 加权图像上,所有情况下都可以看到 3mm 及以上的台阶。20mm 圆的测量值为 19±0.3mm。
所提出的乳腺 MRI 体模可用于获得可重复的测量值,并允许对越来越多地使用的模态实施质量控制措施。